FDA Adverse Event Injury Summary report: N

ENDURANT II STENT GRAFT

MDR report key: 23656912 · Received November 26, 2025

Report

Report Number
9612164-2025-05866
Event Type
Injury
Date Received
November 26, 2025
Date of Event
October 15, 2025
Report Date
November 26, 2025
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC RECEIVED THE FOLLOWING INFORMATION FROM A JOURNAL ARTICLE ENTITLED: ¿ A CASE REPORT OF ACUTE HEMOLYSIS AFTER ENDOVASCULAR ABDOMINAL ANEURYSM REPAIR WITH ONYX EMBOLIZATION OF ENDOLEAK: A CLINICAL CONUNDRUM¿  CHOUDHURY A R, VALAKKADA J, AYAPPAN A, ALEX A JOURNAL OF CLINICAL INTERVENTIONAL RADIOLOGY HTTPS://DOI.ORG/10.1055/S-0045-1812286 DATE OF PUBLISH USED FOR INCIDENT DATE IN SECTION 2.3. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED THE FOLLOWING INFORMATION FROM A JOURNAL ARTICLE ENTITLED ¿ A CASE REPORT OF ACUTE HEMOLYSIS AFTER ENDOVASCULAR ABDOMINAL ANEURYSM REPAIR WITH ONYX EMBOLIZATION OF ENDOLEAK: A CLINICAL CONUNDRUM¿ A PATIENT WAS DIAGNOSED WITH A 75MM FUSIFORM ABDOMINAL AORTIC ANEURYSM. THE ANEURYSM SAC WAS LOCATED 20 MM FROM THE LOWEST RENAL ARTERY (RIGHT RENAL ARTERY). ADDITIONALLY, THERE WAS A SMALL CONTRAST-FILLED OUT-POUCHING SUGGESTIVE OF A PENETRATING ATHEROSCLEROTIC ULCER (PAU) AT THE NECK OF THE ANEURYSM. INTERVENTION WAS PERFORMED AND AN ENDURANT IIS STENT GRAFT SYSTEM WAS IMPLANTED. POST-DEPLOYMENT ANGIOGRAPHY REVEALED ONGOING FILLING OF THE SAC (IA ENDOLEAK) AND PROXIMAL PAU, LIKELY ATTRIBUTED TO INADEQUATE APPOSITION AT THE PROXIMAL LANDING ZONE. THEREFORE THE PAU WAS EMBOLIZED WITH COILS AND AN EMBOLIC AGENT WAS USED IN THE ANEURYSM SAC. THE FINAL ANGIOGRAM INDICATED NO FURTHER ENDOLEAK. OVERALL, THE INTRA-PROCEDURAL BLOOD LOSS WAS LESS THAN 100 ML. ON THE MORNING OF POSTOPERATIVE DAY (POD) 1, THE PATIENT EXHIBITED A SIGNIFICANT DECREASE IN HEMOGLOBIN (HB) LEVELS, DROPPING FROM 11.9 TO 9.1 G/DL. THE PATIENT REMAINED ASYMPTOMATIC AND HEMODYNAMICALLY STABLE, MAINTAINING A BLOOD PRESSURE OF 100/70 MM HG. IMAGING OF THE PATIENTS ABDOMEN CONFIRMED NO ISSUES AND NO RETROPERITONEAL FREE FLUID. HOWEVER BY THAT AFTERNOON THE PATIENTS HB HAD DROPPED TO 8.5G/DL. A CT CONFIRMED NO BLEEDING OR ENDOLEAK. THE PATIENT WAS TRANSFUSED WITH 2 UNITS OF RED BLOOD CELLS. ANOTHER DROP IN HB LEVELS WAS NOTED THE FOLLOWING MORNING ON POST -OP DAY 2 HOWEVER THE PATIENTS HEMOGLOBIN LEVELS STABILISED ON DAY 3 AND NO FURTHER TRANSFUSION WAS REQUIRED. LABORATORY INVESTIGATIONS INDICATED A SIGNIFICANT INCREASE IN INDIRECT BILIRUBIN, RISING FROM A PREPROCEDURAL VALUE OF 0.5 MG/DL TO A POST-PROCEDURAL VALUE OF 1.2 MG/DL, ACCOMPANIED BY AN ISOLATED ELEVATION OF SERUM GLUTAMIC-OXALOACETIC TRANSAMINASE LEVELS. THE POST-PROCEDURE RETICULOCYTE COUNT WAS RECORDED AT 2.5% (ELEVATED) AND AN INCREASED SERUM LACTATE LEVEL,WHICH COLLECTIVELY SUGGESTED ACUTE HEMOLYSIS. THE PATIENT WAS OBSERVED FOR AN ADDITIONAL 2 DAYS AND WAS ULTIMATELY DISCHARGED IN STABLE CONDITION ON POD 5. IT WAS SAID THAT ONE POSSIBLE CAUSE OF THE HEMOLYSIS COULD BE MECHANICAL DAMAGE TO THE RED BLOOD CELLS BY THE BARE METAL PART OF THE STENT AND THE TORTUOSITY OF THE ILIAC VESSELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1946620 ENDURANT II STENT GRAFT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR MIH MEDTRONIC IRELAND UNK-CV-SR-ENDUR-II

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male Required Intervention