COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-02534
- Event Type
- Injury
- Date Received
- December 9, 2011
- Date of Event
- November 17, 2011
- Report Date
- November 17, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS DISPATCHED TO THE SITE FOR THIS EVENT. THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THAT THE ACTUATORS FOR PINCH VALVES VL4 AND VL6 WERE NOT FUNCTIONING PROPERLY. THE ACTUATORS FOR BOTH VALVES WERE REPLACED. THE REPAIR WAS VERIFIED AND SYSTEM VALIDATED PRIOR TO RETURNING THE INSTRUMENT BACK INTO OPERATION. THE ROOT CAUSE OF THE GENERATED ERRONEOUS RESULTS APPEARS TO BE HARDWARE RELATED. THE ROOT CAUSE OF THE REPORTING OF THE ERRONEOUS RESULTS WAS USE ERROR. BECKMAN COULTER INC. LABELING INDICATES THAT THEY SHOULD BE CONTACTED IN THE INSTANCE OF DAILY CHECK FAILURES. MDRS ASSOCIATED WITH THIS EVENT: 1061932-2011-02532, 1061932-2011-02533, 1061932-2011-02534, 1061932-2011-02535.
A CUSTOMER REPORTED THAT A COULTER LH 750 HEMATOLOGY ANALYZER GENERATED ERRONEOUS LOW HEMOGLOBIN (HGB) RESULTS WITH NO INSTRUMENT GENERATED FLAGS FOR MULTIPLE PATIENT SAMPLES OVER TWO DAYS. INSTRUMENT DAILY CHECKS FAILED ON STARTUP FOR HGB DURING THE TIMEFRAME OF THIS EVENT. DAILY CHECKS VERIFY AND CONFIRM IF THE STARTUP TESTS PASS. IF THE DAILY CHECK BACKGROUND FAILS, THEN, PER LABELING, THE CUSTOMER SHOULD CONTACT BECKMAN COULTER. DESPITE FAILING DAILY CHECKS, THE CUSTOMER CONTINUED TO OPERATE AND REPORT OUT PATIENT RESULTS. THIS IS REPORT THREE OF FOUR AND REPRESENTS THE ERRONEOUSLY ELEVATED HGB RESULT GENERATED ON A COULTER LH 750 HEMATOLOGY ANALYZER FOR ONE PATIENT ON (B)(6) 2011. THE SAMPLE WAS RERUN, YIELDING A HIGHER HGB REPEAT RESULT WHICH THE CUSTOMER CONSIDERED CORRECT. THE INITIAL ERRONEOUS HGB RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AND THE PATIENT WAS TRANSFUSED WITH ONE UNIT OF BLOOD BASED UPON THE ERRONEOUS HGB RESULT. NO SPECIFIC PATIENT INFORMATION OR SAMPLE COLLECTION/HANDLING INFORMATION WAS PROVIDED BY THE CUSTOMER. CONTROLS WERE EXECUTED BEFORE THE INCIDENT AND WERE WITHIN THE EXPECTED RANGE. THE INSTRUMENT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS AND REPRODUCIBILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |