FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 23639269 · Received November 24, 2025

Report

Report Number
2955842-2025-46177
Event Type
Malfunction
Date Received
November 24, 2025
Date of Event
October 29, 2025
Report Date
November 24, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874124048
PMA / PMN Number
K241814
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN INSTRUMENT TUBE ADAPTER. THIS COMPONENT IS LOCATED AT THE DISTAL END OF THE INSTRUMENT AND SERVES AS THE INTERFACE BETWEEN THE INSTRUMENT, AND THE USER INSTALLED TIP ACCESSORY. THE BROKEN PIECE WAS NOT RETURNED AND WOULD HAVE MEASURED APPROXIMATELY 5.14MM X 4.46MM IN SIZE. THE BROKEN TIP ADAPTER MADE IT DIFFICULT TO INSTALL A TIP AND CAUSED THE TIP TO NOT BE SECURELY ATTACHED. A VISUAL INSPECTION WAS PERFORMED AND FOUND NO EXTERNAL DAMAGE TO THE SHAFT. THE HOUSING WAS REMOVED FOR INSPECTION AND FOUND NO DAMAGE WITHIN THE INSTRUMENT'S HOUSING. INPUT DISKS ARTICULATED WITH NO ISSUES. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF A DAMAGED TUBE ADAPTER IS ATTRIBUTED TO EITHER TIP ACCESSORY INSTALLATION ISSUES OR DAMAGE DURING USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED NEPHRECTOMY SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT'S TIP WAS BROKEN, AND THE SCISSOR COULD NOT BE PLACED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE TIP WAS NOTICED BEFORE THE CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308467 ENDOWRIST SP MONOPOLAR CURVED SCISSORS ER NAY INTUITIVE SURGICAL, INC 431100-01 K12250110 0019 00886874124048

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES