ELECSYS ANTI-TSHR
Report
- Report Number
- 1823260-2025-04805
- Event Type
- Malfunction
- Date Received
- November 24, 2025
- Date of Event
- August 1, 2025
- Report Date
- December 16, 2025
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JZO
- UDI-DI
- 07613336160305
- PMA / PMN Number
- K193313
- Removal / Correction Number
- RES 98103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ISSUES HAVE BEEN IDENTIFIED WITH THE ELECSYS ANTI-TSHR ASSAY, PARTICULARLY AROUND THE 1.75 IU/L MEDICAL DECISION POINT. ROCHE DIAGNOSTICS HAS ISSUED AN URGENT MEDICAL DEVICE CORRECTION FOR THIS ISSUE, ENTITLED "ELECSYS® ANTI-TSHR ASSAY ¿ DEVIATIONS ON COBAS® E 411, 402, AND 801 ANALYZERS." THIS CORRECTION HAS BEEN REPORTED TO FDA. ROCHE DIAGNOSTICS INVESTIGATED AND CONFIRMED THE CUSTOMERS' ALLEGATIONS. THE ROOT CAUSE FOR THE VARIATIONS IN THE ELECSYS ANTI-TSHR ASSAY WAS AN INSUFFICIENCY OF THE ASSAY'S ROUTINE PRODUCTION PROCESS. FOR THE PLATFORM-TO-PLATFORM DEVIATION (COBAS E 411 VS. COBAS E 402/E 801), DIFFERENCES AROSE DUE TO AN INSUFFICIENT ALIGNMENT CONCEPT BETWEEN THE PLATFORMS. THESE VARIATIONS WERE FURTHER INFLUENCED BY UPDATES TO THE ROUTINE PRODUCTION PROCESS FOR THIS ASSAY. FOR THE LOT-TO-LOT DEVIATION ON E 411, MEASURES TAKEN TO ENHANCE ROBUSTNESS LED TO ADDITIONAL SHIFTS. CUSTOMERS WERE ADVISED TO IMMEDIATELY DISCONTINUE USING AND DISCARD THE AFFECTED LOT NUMBERS: 840183 AND 874011 ON THE COBAS E402/E801 ANALYZER. 840177 ON THE COBAS E411 ANALYZER. CUSTOMERS WERE ADVISED TO SWITCH TO THE NEW UNAFFECTED LOT NUMBERS: 908953 ON THE COBAS E402/E801 ANALYZER. 906872 ON THE COBAS E411 ANALYZER. THE COBAS E601 AND E602 MODULES ARE NOT AFFECTED AND CAN BE USED WITH ALL CURRENTLY AVAILABLE LOTS AND WITH THE UPCOMING LOTS WITHOUT RESTRICTIONS. MEDWATCH FIELDS H7 AND H9 HAVE BEEN UPDATED.
THE SERIAL NUMBER OF THE E411 ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.
THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLY HIGH RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENT SAMPLES TESTED WITH ELECSYS ANTI-TSHR ON A COBAS E411 RACK. THE RESULTS DO NOT AGREE WITH THE CLINICAL CONDITION OF THE PATIENTS. THE SPECIFIC DATE OF THE EVENT IS NOT KNOWN. THE CUSTOMER STATED THEY NOTICED A SHIFT UPWARD IN PATIENT RESULTS SINCE MID (B)(6) 2025 WHEN USING REAGENT LOT 840177 IN COMBINATION WITH CALIBRATOR LOT 849155. THE PATIENT RESULTS RECOVER ABOVE THE NORMAL CUTOFF OF 1.75 IU/L IN PATIENTS WITH AN OTHERWISE EUTHYROID METABOLIC STATE. FOR EXAMPLE, A PATIENT SAMPLE WILL RECOVER A SLIGHTLY ELEVATED VALUE OF 2.2 IU/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2309955 | ELECSYS ANTI-TSHR | THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM | JZO | ROCHE DIAGNOSTICS | 84017701 | 07613336160305 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |