FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-TSHR

MDR report key: 23635053 · Received November 24, 2025

Report

Report Number
1823260-2025-04805
Event Type
Malfunction
Date Received
November 24, 2025
Date of Event
August 1, 2025
Report Date
December 16, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JZO
UDI-DI
07613336160305
PMA / PMN Number
K193313
Removal / Correction Number
RES 98103
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ISSUES HAVE BEEN IDENTIFIED WITH THE ELECSYS ANTI-TSHR ASSAY, PARTICULARLY AROUND THE 1.75 IU/L MEDICAL DECISION POINT. ROCHE DIAGNOSTICS HAS ISSUED AN URGENT MEDICAL DEVICE CORRECTION FOR THIS ISSUE, ENTITLED "ELECSYS® ANTI-TSHR ASSAY ¿ DEVIATIONS ON COBAS® E 411, 402, AND 801 ANALYZERS." THIS CORRECTION HAS BEEN REPORTED TO FDA. ROCHE DIAGNOSTICS INVESTIGATED AND CONFIRMED THE CUSTOMERS' ALLEGATIONS. THE ROOT CAUSE FOR THE VARIATIONS IN THE ELECSYS ANTI-TSHR ASSAY WAS AN INSUFFICIENCY OF THE ASSAY'S ROUTINE PRODUCTION PROCESS. FOR THE PLATFORM-TO-PLATFORM DEVIATION (COBAS E 411 VS. COBAS E 402/E 801), DIFFERENCES AROSE DUE TO AN INSUFFICIENT ALIGNMENT CONCEPT BETWEEN THE PLATFORMS. THESE VARIATIONS WERE FURTHER INFLUENCED BY UPDATES TO THE ROUTINE PRODUCTION PROCESS FOR THIS ASSAY. FOR THE LOT-TO-LOT DEVIATION ON E 411, MEASURES TAKEN TO ENHANCE ROBUSTNESS LED TO ADDITIONAL SHIFTS. CUSTOMERS WERE ADVISED TO IMMEDIATELY DISCONTINUE USING AND DISCARD THE AFFECTED LOT NUMBERS: 840183 AND 874011 ON THE COBAS E402/E801 ANALYZER. 840177 ON THE COBAS E411 ANALYZER. CUSTOMERS WERE ADVISED TO SWITCH TO THE NEW UNAFFECTED LOT NUMBERS: 908953 ON THE COBAS E402/E801 ANALYZER. 906872 ON THE COBAS E411 ANALYZER. THE COBAS E601 AND E602 MODULES ARE NOT AFFECTED AND CAN BE USED WITH ALL CURRENTLY AVAILABLE LOTS AND WITH THE UPCOMING LOTS WITHOUT RESTRICTIONS. MEDWATCH FIELDS H7 AND H9 HAVE BEEN UPDATED.

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE E411 ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLY HIGH RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENT SAMPLES TESTED WITH ELECSYS ANTI-TSHR ON A COBAS E411 RACK. THE RESULTS DO NOT AGREE WITH THE CLINICAL CONDITION OF THE PATIENTS. THE SPECIFIC DATE OF THE EVENT IS NOT KNOWN. THE CUSTOMER STATED THEY NOTICED A SHIFT UPWARD IN PATIENT RESULTS SINCE MID (B)(6) 2025 WHEN USING REAGENT LOT 840177 IN COMBINATION WITH CALIBRATOR LOT 849155. THE PATIENT RESULTS RECOVER ABOVE THE NORMAL CUTOFF OF 1.75 IU/L IN PATIENTS WITH AN OTHERWISE EUTHYROID METABOLIC STATE. FOR EXAMPLE, A PATIENT SAMPLE WILL RECOVER A SLIGHTLY ELEVATED VALUE OF 2.2 IU/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2309955 ELECSYS ANTI-TSHR THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM JZO ROCHE DIAGNOSTICS 84017701 07613336160305

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown