PCCO MONITORING KIT

MDR report key: 23630166 · Received November 24, 2025

Report

Report Number
3003263092-2025-000002
Event Type
Malfunction
Date Received
November 24, 2025
Date of Event
October 26, 2025
Report Date
November 24, 2025
Manufacturer
PULSION MEDICAL SYSTEMS SE
Product Code
KRB
UDI-DI
04250094500757
PMA / PMN Number
K991886
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT A CRACK IN STOP-COCK CONNECTOR OCCURRED, RESULTING IN BLOOD LEAKAGE FROM THE DISCONNECTED PORT. NO PATIENT HARM HAS BEEN REPORTED. THE FAULTY PART WAS DISCARDED BY THE CUSTOMER, THE PICTURE OF BROKEN PART WAS PROVIDED, SO THE ISSUE COULD BE CONFIRMED. THE CAUSE OF THE ISSUE COULD NOT BE DETERMINED. THERE IS NO INDICATION FOR A SYSTEMATIC ROOT CAUSE AS A DEFICIENCY OF DESIGN, PRODUCTION OR MATERIAL. OVERALL, INVESTIGATIONS DID INDICATE THAT THE DEVICE FAILED TO MEET ITS SPECIFICATION WHEN THE EVENT OCCURRED. A RELATIONSHIP BETWEEN THE DEVICE AND THE COMPLAINT IS THEREFORE CONSIDERED AS LIKELY. UPON THE COMPLAINT OCCURRENCE THE DEVICE WAS INVOLVED AND USED ON A PATIENT FOR ADVANCED HEMODYNAMIC MONITORING. IN GENERAL A MONITORING TECHNOLOGY IS CONSIDERED AS DIAGNOSTIC AID, BUT NOT DIRECTLY USED FOR DIAGNOSIS OR TREATMENT. THIS EVENT OCCURRED ON THE CHINESE MARKET ON A DEVICE THAT IS DISTRIBUTED EXCLUSIVELY OUTSIDE OF THE USA. THEREFORE, NO GUDID INFORMATION EXISTS. UDI INFORMATION PROVIDED IN D4 IS FOR THE OUS DEVICE. IT IS A SIGNIFICANTLY SIMILAR DEVICE TO ¿PICCO MONITORING KIT¿ WHICH IS SOLD IN THE USA UNDER PRIMARY DI NUMBER (B)(4), PREMARKET SUBMISSION NUMBER K991886.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PART OF PICCO MONITORING KIT HAS DEVELOPED A CRACK, RESULTING IN BLOOD LEAKAGE FROM THE DISCONNECTED PORT. THIS EVENT DID NOT CAUSE ANY ADDITIONAL HARM TO THE PATIENT. MANUFACTURER'S REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2749799 PCCO MONITORING KIT PROBE, THERMODILUTION KRB PULSION MEDICAL SYSTEMS SE PV8215 24AK26 04250094500757

Patients

Seq Age Sex Outcome Treatment
1 78 YR Female