FDA Adverse Event
Injury
Summary report: N
PRIMARY MICROBORE GEMSTAR SET WITH 0.2MICRON FILTER
MDR report key: 2362747
·
Received December 1, 2011
Report
- Report Number
- MW5023335
- Event Type
- Injury
- Date Received
- December 1, 2011
- Date of Event
- November 24, 2011
- Report Date
- December 1, 2011
- Product Code
- FPA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MI, US
- Reporter Occupation
- PHARMACIST
Narratives
Description of Event or Problem · 1
WE HAVE HAD TWO PATIENTS ON CONTINUOUS IV PIPERCILLIN TAZOBACTAM ON A HOSPIRA GEMSTAR PUMP. ON THREE OCCASIONS IN THE LAST TWO WEEKS, I HAVE HAD PATIENTS CALL COMPLAINING OF AN OCCLUSION ALARM. AFTER SWITCHING INFUSION TUBING THE OCCLUSION ALARM GOES AWAY TELLING ME WE ARE MOST LIKELY HAVING DRUG PRECIPITATE IN THE FILTER CAUSING AND OCCLUSION. DOSE: 18GM OVER 24 HOURS. DATES OF USE: (B)(6) 2011 TO PRESENT. REASON FOR USE: DIABETIC FOOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRIMARY MICROBORE GEMSTAR SET WITH 0.2MICRON FILTER | GEMSTAR 0.2 MICRON SET | FPA | 13759-28 | 07175 5H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | PIPERACILLIN TAZOBACTAM 40.5GM BULK,| LOT 503B5020, EXP DATE 07/2013| APOTEX CORP/HOSPIRA |