FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® Z BLOOD COLLECTION TUBES

MDR report key: 23618129 · Received November 21, 2025

Report

Report Number
9617032-2025-02076
Event Type
Malfunction
Date Received
November 21, 2025
Date of Event
October 30, 2025
Report Date
January 22, 2026
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
30382903676140
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D4. MEDICAL DEVICE LOT#: 5240041 D4. MEDICAL DEVICE EXPIRATION DATE: 28-FEB-2027 D4. UNIQUE IDENTIFIER (UDI) #: (B)(4). H4. DEVICE MANUFACTURE DATE: 28-AUG-2025 INVESTIGATION SUMMARY: BD RECEIVED ONE PHOTO FOR INVESTIGATION. THE EVALUATION OF THE PHOTOGRAPH SHOWCASES A SHELF PACK LABEL OF THE PRODUCT. ADDITIVE ABNORMALITY COULD NOT BE SEEN IN THE PHOTO. ADDITIONALLY, 100 RETAINED SAMPLES WERE VISUALLY EXAMINED, AND NO ADDITIVE ABNORMALITIES WERE FOUND. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ADDITIVE ABNORMALITY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED ONE PHOTO FOR INVESTIGATION. THE EVALUATION OF THE PHOTOGRAPH SHOWCASES A SHELF PACK LABEL OF THE PRODUCT. ADDITIVE ABNORMALITY COULD NOT BE SEEN IN THE PHOTO. ADDITIONALLY, 100 RETAINED SAMPLES WERE VISUALLY EXAMINED, AND NO ADDITIVE ABNORMALITIES WERE FOUND. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ADDITIVE ABNORMALITY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

B3: DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. D4. MEDICAL DEVICE LOT#: UNKNOWN. D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® Z BLOOD COLLECTION TUBES, LIQUID, ABNORMAL ADDITIVE, IS SEEN INSIDE OF AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® Z BLOOD COLLECTION TUBES, LIQUID, ABNORMAL ADDITIVE, IS SEEN INSIDE OF AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® Z BLOOD COLLECTION TUBES, LIQUID, ABNORMAL ADDITIVE, IS SEEN INSIDE OF AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2463362 BD VACUTAINER® Z BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 5240041 30382903676140

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown