FDA Adverse Event Malfunction Summary report: N

DEXTER

MDR report key: 23616367 · Received November 21, 2025

Report

Report Number
3033118964-2025-00002
Event Type
Malfunction
Date Received
November 21, 2025
Date of Event
October 23, 2025
Report Date
November 21, 2025
Manufacturer
DISTALMOTION SA
Product Code
SDD
UDI-DI
07640184840622
PMA / PMN Number
K251197
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE MONOPOLAR HOOK INSTRUMENT WAS RETURNED FOR ANALYSIS. INSPECTION CONFIRMED INSULATION DAMAGE IN THE TIP SLEEVE REGION; TWO HOLE WERE OBSERVED. REVIEW OF PRODUCTION RECORDS REVEALED NO MANUFACTURING ANOMALIES, AND NO EVIDENCE OF MATERIAL OR ASSEMBLY DEFECT WAS FOUND. THE FIRST HOLE WAS LIKELY CAUSED BY CONTACT WITH THE DISTAL FORK UNDER HIGH ANGULATION OR PROXIMITY TO OTHER DEVICES DURING USE. THIS DAMAGE CAN COMPROMISE INSULATION, LEADING TO SHORT CIRCUIT AND THERMAL DAMAGE WHEN ENERGIZED. THE DAMAGE IS CONSISTENT WITH MECHANICAL STRESS DURING USE; HOWEVER, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. THE SECOND HOLE WAS LIKELY DAMAGED BY EXTERNAL INTERACTION. THE DAMAGED INSULATION IN CONTACT WITH TISSUE EVENTUALLY LEADS TO SHORT CIRCUIT WHEN THE TIP OF THE MONOPOLAR INSTRUMENT IS ENERGIZED.

Description of Event or Problem · 0

DURING TISSUE COAGULATION IN A REFLUXSTOP¿ DEVICE IMPLANTATION PROCEDURE, IT WAS REPORTED THAT AFTER APPROXIMATELY 40 MINUTES OF USING THE MONOPOLAR HOOK, A SPARK TRAIL INDICATED A DEFECT IN THE TIP SLEEVE. THERE WAS NO HARM TO THE PATIENT. THE INSTRUMENT WAS IMMEDIATELY REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND WITHOUT INCIDENT. THERE WERE NO REPORTS OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2117399 DEXTER DEXTER MONOPOLAR HOOK SDD DISTALMOTION SA 520 520A202502X11546 07640184840622

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female