FDA Adverse Event Injury Summary report: N

BIPOLAR HEADS

MDR report key: 23615344 · Received November 21, 2025

Report

Report Number
3005180920-2025-01165
Event Type
Injury
Date Received
November 21, 2025
Date of Event
November 6, 2025
Report Date
November 21, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
KWY
UDI-DI
07630030843679
PMA / PMN Number
K091967
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 11 NOV 2025. BALL HEADS: BIPOLAR HEAD 25060.2848 COCR BIPOLAR HEAD D 28X48 (K091967) LOT. 2410246: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-SEPT-2024. EXPIRATION DATE: 25-AUG-2029. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE NO DEFINITIVE ROOT CAUSE CAN BE ESTABLISHED, WHILE THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

AFTER TWO MONTHS FROM THE PRIMARY SURGERY, THE BIPOLAR HEAD DISLOCATED FROM THE ACETABULUM. THE SURGEON REVISED THE IMPLANT, CONVERTING THE BIPOLAR SYSTEM TO A TOTAL HIP ARTHROPLASTY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2024013 BIPOLAR HEADS COCR BIPOLAR HEAD Ø28X48 KWY MEDACTA INTERNATIONAL SA 25060.2848 2410246 07630030843679

Patients

Seq Age Sex Outcome Treatment
1 86 YR Female Required Intervention