FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ NMIC/ID-307

MDR report key: 23609704 · Received November 20, 2025

Report

Report Number
1119779-2025-05448
Event Type
Malfunction
Date Received
November 20, 2025
Date of Event
October 20, 2025
Report Date
March 31, 2026
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904492893
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC RESULTS FOR IMIPENEM (IPM) AND ERTAPENEM (ETP) WHEN USING PANEL PHOENIX NMIC/ID-307 (CATALOG NUMBER 449289) BATCH NUMBER 5189220. THE CUSTOMER RETURNED PHOENIX GENERATED LAB REPORTS WITH RESULTS WITH ESCHERICHIA COLI FOR THE INVESTIGATION. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS ANALYZED, AND NO CURRENT TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. HISTORICAL TRENDING WAS REVIEWED AND THERE WERE NO TRENDS FOR HIGH MIC RESULTS FOR PANEL SKU 449289. TO INVESTIGATE, RETENTION PANELS FROM THE COMPLAINT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES ESCHERICHIA COLI A25922 AND ESCHERICHIA COLI 11002 TO OBSERVE FOR IPM AND ETP MIC RESULTS. NEXT, CONTROL PANELS FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES ESCHERICHIA COLI A25922 AND ESCHERICHIA COLI 11002 TO OBSERVE FOR IPM AND ETP MIC RESULTS. THE INVESTIGATION RETURNED ALL PANELS WITH SATISFACTORY SUSCEPTIBLE IPM AND ETP SIR AND MIC RESULTS, THEREFORE THIS COMPLAINT IS NOT CONFIRMED. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA /510(K)#: G4. PMA / 510(K)# K020322, K023444, K023634, K023858, K024153, K031530, K031699, K032299, K032655, K033560, K041384, K042932, K052269, K060214, K060217, K060444, K060447, K061355, K062944, K063301, K063573, K063811, K063824, K071623, K132674, K151320, K181665, K033458 AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
111233 BD PHOENIX¿ NMIC/ID-307 SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) 5189220 30382904492893

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown