BD PHOENIX¿ NMIC/ID-307
Report
- Report Number
- 1119779-2025-05448
- Event Type
- Malfunction
- Date Received
- November 20, 2025
- Date of Event
- October 20, 2025
- Report Date
- March 31, 2026
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904492893
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC RESULTS FOR IMIPENEM (IPM) AND ERTAPENEM (ETP) WHEN USING PANEL PHOENIX NMIC/ID-307 (CATALOG NUMBER 449289) BATCH NUMBER 5189220. THE CUSTOMER RETURNED PHOENIX GENERATED LAB REPORTS WITH RESULTS WITH ESCHERICHIA COLI FOR THE INVESTIGATION. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS ANALYZED, AND NO CURRENT TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. HISTORICAL TRENDING WAS REVIEWED AND THERE WERE NO TRENDS FOR HIGH MIC RESULTS FOR PANEL SKU 449289. TO INVESTIGATE, RETENTION PANELS FROM THE COMPLAINT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES ESCHERICHIA COLI A25922 AND ESCHERICHIA COLI 11002 TO OBSERVE FOR IPM AND ETP MIC RESULTS. NEXT, CONTROL PANELS FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES ESCHERICHIA COLI A25922 AND ESCHERICHIA COLI 11002 TO OBSERVE FOR IPM AND ETP MIC RESULTS. THE INVESTIGATION RETURNED ALL PANELS WITH SATISFACTORY SUSCEPTIBLE IPM AND ETP SIR AND MIC RESULTS, THEREFORE THIS COMPLAINT IS NOT CONFIRMED. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA /510(K)#: G4. PMA / 510(K)# K020322, K023444, K023634, K023858, K024153, K031530, K031699, K032299, K032655, K033560, K041384, K042932, K052269, K060214, K060217, K060444, K060447, K061355, K062944, K063301, K063573, K063811, K063824, K071623, K132674, K151320, K181665, K033458 AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 111233 | BD PHOENIX¿ NMIC/ID-307 | SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION | LON | BECTON DICKINSON & CO. (SPARKS) | 5189220 | 30382904492893 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |