QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8
Report
- Report Number
- 3016746283-2025-00017
- Event Type
- Injury
- Date Received
- November 20, 2025
- Date of Event
- October 24, 2025
- Report Date
- May 11, 2026
- Manufacturer
- SPECTRUM MEDICAL S.R.L.
- Product Code
- DTR
- UDI-DI
- 08051160300501
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION RELATED TO THE EVENT WILL BE AVAILABLE AFTER COMPLETION OF INVESTIGATION ACTIVITIES, WHICH WOULD BE ADDRESSED IN A FOLLOW-UP REPORT.
REFER TO QHCX0501_INVESTIGATION_RMD_PRELIMINARY FOR PARTIAL CONCLUSIONS.
VISUAL INSPECTION AND DEFECT IDENTIFICATION: TWO SYMMETRICAL HOLES WERE FOUND IN THE CENTRAL AREA OF THE DEVICE WHERE THE KNURLED PATTERN CHANGES DIRECTION. SCANNING ELECTRON MICROSCOPY (SEM): SEM IMAGING WAS USED TO CHARACTERIZE THE DIMENSIONS OF THE HOLES (APPROXIMATELY 100 MICRONS) AND TO OBTAIN DETAILS OF THE METAL SHEET SURFACE NEAR THE HOLES. FLUID TEST RESULTS: TESTING CONFIRMED THAT EVEN WITH THE HOLES PRESENT, UNDER WORST-CASE CONDITIONS, FLOW OCCURS FROM BLOOD TO HTF, ENSURING THAT NO HTF CAN ENTER THE BLOOD COMPARTMENT. THIS IS DUE TO THE FROSTY DESIGN: HTF IS DRAWN OUT OF THE HEAT EXCHANGER, SO THE PRESSURE INSIDE THE HTF COMPARTMENT IS ALWAYS NEGATIVE, WHILE THE BLOOD COMPARTMENT IS ALWAYS UNDER POSITIVE PRESSURE. AS A RESULT, THE PRESSURE DIFFERENTIAL ACROSS THE METAL SHEET GUARANTEES THAT ONLY BLOOD CAN ENTER THE HTF COMPARTMENT, NOT VICE VERSA. A FINAL AND COMPLETE REPORT WILL BE RELEASED WHEN ALL THE RESULTS WILL BE AVAILABLE.
AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.
AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.
AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176412 | QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8 | STANDARD HEAT EXCHANGER | DTR | SPECTRUM MEDICAL S.R.L. | HX55V-S0 | G002192 | 08051160300501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |