FDA Adverse Event Injury Summary report: N

QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8

MDR report key: 23603338 · Received November 20, 2025

Report

Report Number
3016746283-2025-00017
Event Type
Injury
Date Received
November 20, 2025
Date of Event
October 24, 2025
Report Date
May 11, 2026
Manufacturer
SPECTRUM MEDICAL S.R.L.
Product Code
DTR
UDI-DI
08051160300501
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RELATED TO THE EVENT WILL BE AVAILABLE AFTER COMPLETION OF INVESTIGATION ACTIVITIES, WHICH WOULD BE ADDRESSED IN A FOLLOW-UP REPORT.

Additional Manufacturer Narrative · 0

REFER TO QHCX0501_INVESTIGATION_RMD_PRELIMINARY FOR PARTIAL CONCLUSIONS.

Additional Manufacturer Narrative · 0

VISUAL INSPECTION AND DEFECT IDENTIFICATION: TWO SYMMETRICAL HOLES WERE FOUND IN THE CENTRAL AREA OF THE DEVICE WHERE THE KNURLED PATTERN CHANGES DIRECTION. SCANNING ELECTRON MICROSCOPY (SEM): SEM IMAGING WAS USED TO CHARACTERIZE THE DIMENSIONS OF THE HOLES (APPROXIMATELY 100 MICRONS) AND TO OBTAIN DETAILS OF THE METAL SHEET SURFACE NEAR THE HOLES. FLUID TEST RESULTS: TESTING CONFIRMED THAT EVEN WITH THE HOLES PRESENT, UNDER WORST-CASE CONDITIONS, FLOW OCCURS FROM BLOOD TO HTF, ENSURING THAT NO HTF CAN ENTER THE BLOOD COMPARTMENT. THIS IS DUE TO THE FROSTY DESIGN: HTF IS DRAWN OUT OF THE HEAT EXCHANGER, SO THE PRESSURE INSIDE THE HTF COMPARTMENT IS ALWAYS NEGATIVE, WHILE THE BLOOD COMPARTMENT IS ALWAYS UNDER POSITIVE PRESSURE. AS A RESULT, THE PRESSURE DIFFERENTIAL ACROSS THE METAL SHEET GUARANTEES THAT ONLY BLOOD CAN ENTER THE HTF COMPARTMENT, NOT VICE VERSA. A FINAL AND COMPLETE REPORT WILL BE RELEASED WHEN ALL THE RESULTS WILL BE AVAILABLE.

Description of Event or Problem · 0

AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.

Description of Event or Problem · 0

AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.

Description of Event or Problem · 0

AFTER CPB AND DURING DEPRIMMING THE CUSTOMER NOTICED THAT THE GLYCOL HAD TURNED IN GREY. BECAUSE THIS EVENT HAPPENED END OF THE CASE, PRODUCT WERE NOT REPLACED. CONSEQUENCES: RESIDUAL BLOOD HAS BEEN WASHED BY CELLSAVER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176412 QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8 STANDARD HEAT EXCHANGER DTR SPECTRUM MEDICAL S.R.L. HX55V-S0 G002192 08051160300501

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention