INSET 30
Report
- Report Number
- 3003442380-2025-16790
- Event Type
- Injury
- Date Received
- November 20, 2025
- Date of Event
- November 2, 2025
- Report Date
- November 22, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011186, IN QUESTION WAS MANUFACTURED AT THE: REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 19-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6011186". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011186 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINE 10 ON 29-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. GLUING OF TUBING OF THE LOT 5A03588 MANUFACTURED IN THE MACHINE ITL03, ON 21-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED, THEREFORE, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH: WI GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3: 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2: 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST FOR THE CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO DEFECT ON TEST OF REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CAPA DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NC (B)(4) 1. INCLUDE THE DEFECT IN DOCUMENT (B)(4) (WORK INSTRUCTION INSET LINE) 2. IMPROVE GUARDS OF THE CONVEYORS 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS 4. UPDATE DOCUMENT (B)(4) (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. 5. UPDATE THE DOCUMENT (B)(4) (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS) A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (DATABASE (B)(4) - (B)(6) 2025).
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED KINKED CANNULA EVENT AND WENT TO THE EMERGENCY ROOM (ER) ON (B)(6) 2025.THE EVENT OCCURRED WITHIN THREE OR MORE HOURS AFTER INSERTION. THE INSERTION SITE WAS ABDOMEN.THE PATIENT WAS IN EMERGENCY ROOM FOR THREE HOURS.THE BLOOD GLUCOSE LEVEL WAS 450 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND INSULIN.THE PATIENT WAS POSITIVE FOR KETONES.THE PATIENT WAS RELEASED FROM THE HOSPITAL ON (B)(6) 2025.THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309607 | INSET 30 | UNO INSET 30 60/13 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001680 | 6011186 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Female | Required Intervention |