FDA Adverse Event Malfunction Summary report: N

FIBULINK® SYNDESMOSIS REPAIR KIT/ TI

MDR report key: 23598415 · Received November 19, 2025

Report

Report Number
8030965-2025-11611
Event Type
Malfunction
Date Received
November 19, 2025
Date of Event
November 4, 2025
Manufacturer
SYNTHES GMBH
Product Code
HTN
UDI-DI
00863176000317
PMA / PMN Number
K162805
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11. ADDITIONAL NARRATIVE: H6. COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. INVESTIGATION SUMMARY: PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: PART# FGS-1100. LOT # 24E019. RELEASE TO WAREHOUSE DATE: 17 DEC 2024. EXPIRATION DATE: 01 DEC 2027. MANUFACTURING SITE: ROBLING MEDICAL, INC. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT ORIF SURGERY WITH FIBULINK FOR FIBULA FRACTURES (WEBER-C). WHILE TENSIONING THE FIBULINK IMPLANT, THE TENSION SUDDENLY DISAPPEARED. THE IMPLANT WAS REMOVED, AND IT WAS FOUND THAT THE PERMACORD HAD BROKEN. THE SURGEON WAS HOLDING THE SILVER TUBE WITH PLIERS AND ROTATING THE TENSION CAP CLOCKWISE, BUT IT APPEARS THAT THE GRIP ON THE SILVER TUBE WAS WEAK, CAUSING THE PERMACORD TO TWIST AND BREAK. SINCE THE BONE QUALITY WAS GOOD, THE PROCEDURE WAS COMPLETED WITHOUT FIBULINK. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2388752 FIBULINK® SYNDESMOSIS REPAIR KIT/ TI WASHER, BOLT NUT HTN SYNTHES GMBH 24E019 00863176000317

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male