FDA Adverse Event Malfunction Summary report: N

ALLURA XPER FD

MDR report key: 23598163 · Received November 19, 2025

Report

Report Number
3003768277-2025-014966
Event Type
Malfunction
Date Received
November 19, 2025
Date of Event
October 30, 2025
Report Date
January 16, 2026
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product Code
IZI
PMA / PMN Number
K130842
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PHILIPS HAS INVESTIGATED THIS COMPLAINT. ACCORDING TO THE ADDITIONAL INFORMATION COLLECTED, THE ISSUE OCCURRED DURING PERFORMING CORONARY PLANNED TREATMENT. THE PROCEDURE WAS DELAYED FOR ONE HOUR AND IT WAS COMPLETED BY USING ANOTHER SYSTEM. THE PHILIPS REMOTE SERVICE ENGINEER (RSE) ANALYZED THE SYSTEM REMOTELY AND CONFIRMED THAT THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY. REVIEW OF THE LOGFILE SHOWS FD-CONTROLLER: IOBOARD UART PARITY ERROR CONFIRMING THE MALFUNCTION. UPON TROUBLESHOOTING, THE PHILIPS FIELD SERVICE ENGINEER (FSE) CHECKED THE SYSTEM ONSITE AND FOUND THAT THE VOLTAGE WAS UNSTABLE AND IT HAD DROPPED IN THE UNIT THAT SUPPLIES POWER TO THE FPD (DETECTOR). FSE CONFIRMED THE RELAY FOR THE FPD POWER SUPPLY IN THE M CABINET WAS NOT OPERATING. TO RESOLVE THE ISSUE FSE REPLACED THE POWER SUPPLY FLAT DETECTOR (PS FD) UNIT. THE DEFECTIVE PART WAS SENT FOR ANALYSIS, FOR FD CONTROLLER FAILURE WAS OBSERVED AND THE FAILED COMPONENT WAS FOUND TO BE FAULTY CAPACITORS. AFTER REPLACEMENT, THE SYSTEM RETURNED TO USE IN GOOD WORKING ORDER. THE CODES WERE UPDATED BASED ON THE INVESTIGATION OUTCOME.

Description of Event or Problem · 0

IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY.THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS.PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2310247 ALLURA XPER FD SYSTEM, X-RAY, ANGIOGRAPHIC IZI PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD20

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown