FDA Adverse Event Malfunction Summary report: N

MILLEX -GV

MDR report key: 23597509 · Received November 19, 2025

Report

Report Number
MW5179067
Event Type
Malfunction
Date Received
November 19, 2025
Date of Event
November 14, 2025
Report Date
November 14, 2025
Manufacturer
MERK MILLIPORE LTD.
Product Code
BSN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Description of Event or Problem · 0

FILTER FAILED INTEGRITY TESTING WITH BUBBLE POINT BETWEEN 35-45 PSI INSTEAD OF LABELED 50 PSI.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2023813 MILLEX -GV FILTER, CONDUCTION, ANESTHETIC BSN MERK MILLIPORE LTD. SLGVM33RS 0000340077

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown