FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD

MDR report key: 23589364 · Received November 18, 2025

Report

Report Number
1710034-2025-01814
Event Type
Malfunction
Date Received
November 18, 2025
Date of Event
October 29, 2025
Report Date
December 16, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825448
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382544 AND LOT NUMBER 5120110 AND 5052302. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

THESE THREE OCCURRED ON 10/29. ALL THREE THE NEEDLE WAS SLOW TO RETRACT.

Description of Event or Problem · 0

NO NEW INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1767821 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5120110 00382903825448

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown