FDA Adverse Event Injury Summary report: N

INSET 30

MDR report key: 23580072 · Received November 18, 2025

Report

Report Number
3003442380-2025-16691
Event Type
Injury
Date Received
November 18, 2025
Date of Event
October 28, 2025
Report Date
November 18, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018365
PMA / PMN Number
K061374
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6013265, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 14-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL #6013265. THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6013265 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 30 AND MANUFACTURED IN THE INSET 30 LI ON 18-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE BAG SEALING, LOT 5E00975 WAS MANUFACTURED ACCORDING TO THE WI VERSION 30 AND MANUFACTURED IN THE LI 3010, ON 15-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5D05151 WAS MANUFACTURED ACCORDING TO THE WI VERSION 67 AND MANUFACTURED IN THE SC07, ON 10-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED. IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE(S) FROM THE LOT HAVE BEEN REQUESTED. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH: WI GUIDANCE FOR VISUAL TEST FOR COMPLAINTS AREA VERSION 3: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. WI GUIDANCE FOR FUNCTIONAL TESTING 1 AIR FLOW TEST FOR COMPLAINTS AREA VERSION 2: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. WI GUIDANCE FOR FUNCTIONAL TESTING 2 AIR LEAK TEST FOR COMPLAINTS AREA VERSION 2: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, 1 COMPLAINT IN THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO HYPERGLYCEMIA ON (B)(6) 2025. THE BLOOD GLUCOSE LEVELS AT THE TIME OF EVENT WAS 500MG/DL. THE CAUSE OF HYPERGLYCEMIA WAS BENT NEEDLE. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1783401 INSET 30 UNO INSET 30 60/13 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002825 6013265 05705244018365

Patients

Seq Age Sex Outcome Treatment
1 43 YR Female Hospitalization| L