IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM
Report
- Report Number
- 0001038806-2025-03222
- Event Type
- Injury
- Date Received
- November 17, 2025
- Date of Event
- October 15, 2025
- Report Date
- November 17, 2025
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- UDI-DI
- 00889024019515
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER: (B)(4). G4: ADDITIONAL PMA/510(K) NUMBER ¿ K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCE'S AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NON-EXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT FIVE (5) IMPLANTS WERE PLACED FOR MAXILLARY PROTHESIS. THE IMPLANT AT TOOTH SITE #4 DEVELOPED DELAYED ABSCESS AFTER A NUMBER OF MONTHS UNDER FULL DENTURE. THE IMPLANT WAS SURGICALLY REMOVED DUE TO PERI-IMPLANTITIS. ADDITIONAL RIDGE AUGMENTATION/GRAFTING WAS ALSO PERFORMED. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS, EDEMA AND INFLAMMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1810692 | IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 1293173 | 00889024019515 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Male | Required Intervention |