BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE
Report
- Report Number
- 1024879-2025-01818
- Event Type
- Malfunction
- Date Received
- November 17, 2025
- Date of Event
- October 20, 2025
- Report Date
- March 23, 2026
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- UDI-DI
- 30382903686071
- PMA / PMN Number
- K982541
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D2A. MEDICAL DEVICE TYPE: JKA. G.4. PMA / 510(K)#:K243207. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED ANY SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, A TOTAL OF 20 RETAINED SAMPLES WERE SUBJECTED TO FUNCTIONAL TESTS FOR DEFECTIVE LOCKING MECHANISM AND HUB COLLAR SEPARATION, WITH NO DEFECTS OBSERVED AS ALL SAMPLES PASSED THE TESTS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: HUB/COLLAR SEPARATION. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BEFORE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, HUB/COLLAR SEPARATION OCCURRED ON AN UNSPECIFIED NUMBER OF UNITS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
IT WAS REPORTED THAT BEFORE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, HUB/COLLAR SEPARATION OCCURRED ON AN UNSPECIFIED NUMBER OF UNITS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 32602 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON, DICKINSON & CO., (BD) | 5134077 | 30382903686071 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |