ALINITY I TOTAL B-HCG REAGENT KIT
Report
- Report Number
- 3005094123-2025-00576
- Event Type
- Malfunction
- Date Received
- November 17, 2025
- Date of Event
- November 5, 2025
- Report Date
- February 4, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- DHA
- UDI-DI
- 00380740163334
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07P51-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P51-21 / 31, WITH 510K/PMA/BLA NUMBER K170317.
THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY I TOTAL SS-HCG RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ALINITY I TOTAL SS-HCG ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 71475UD01. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES, OR DEVIATIONS ASSOCIATED WITH THE LOT NUMBER AND COMPLAINT ISSUE. THE OVERALL PERFORMANCE OF ALINITY I TOTAL SS-HCG REAGENT WAS REVIEWED USING WORLDWIDE FIELD DATA. REVIEW SHOWS THAT THE MEDIAN PATIENT RESULT FOR THE LOT IS WITHIN ESTABLISHED LIMITS AND COMPARABLE WITH OTHER LOTS IN THE FIELD, CONFIRMING NO SYSTEMIC ISSUE FOR THE LOT. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I TOTAL SS-HCG REAGENT LOT 71475UD01 WAS IDENTIFIED. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I TOTAL -HCG RESULT FOR A NON-PREGNANT 21-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED REFERENCE RANGE 0-5 MIU/ML) INITIAL RESULT = 40.91 MIU/ML, REPEAT RESULT = <1.1 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I TOTAL -HCG RESULT FOR A NON-PREGNANT 21-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED REFERENCE RANGE 0-5 MIU/ML) INITIAL RESULT = 40.91 MIU/ML, REPEAT RESULT = <1.1 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2881690 | ALINITY I TOTAL B-HCG REAGENT KIT | SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN | DHA | ABBOTT IRELAND DIAGNOSTICS DIVISION | 71475UD01 | 00380740163334 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Female | ALNTY I PROCESSING MODU, 03R65-01,(B)(6) |