FDA Adverse Event Malfunction Summary report: N

ALINITY I TOTAL B-HCG REAGENT KIT

MDR report key: 23568768 · Received November 17, 2025

Report

Report Number
3005094123-2025-00576
Event Type
Malfunction
Date Received
November 17, 2025
Date of Event
November 5, 2025
Report Date
February 4, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
DHA
UDI-DI
00380740163334
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07P51-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P51-21 / 31, WITH 510K/PMA/BLA NUMBER K170317.

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY I TOTAL SS-HCG RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ALINITY I TOTAL SS-HCG ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 71475UD01. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES, OR DEVIATIONS ASSOCIATED WITH THE LOT NUMBER AND COMPLAINT ISSUE. THE OVERALL PERFORMANCE OF ALINITY I TOTAL SS-HCG REAGENT WAS REVIEWED USING WORLDWIDE FIELD DATA. REVIEW SHOWS THAT THE MEDIAN PATIENT RESULT FOR THE LOT IS WITHIN ESTABLISHED LIMITS AND COMPARABLE WITH OTHER LOTS IN THE FIELD, CONFIRMING NO SYSTEMIC ISSUE FOR THE LOT. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I TOTAL SS-HCG REAGENT LOT 71475UD01 WAS IDENTIFIED. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I TOTAL -HCG RESULT FOR A NON-PREGNANT 21-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED REFERENCE RANGE 0-5 MIU/ML) INITIAL RESULT = 40.91 MIU/ML, REPEAT RESULT = <1.1 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I TOTAL -HCG RESULT FOR A NON-PREGNANT 21-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED REFERENCE RANGE 0-5 MIU/ML) INITIAL RESULT = 40.91 MIU/ML, REPEAT RESULT = <1.1 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2881690 ALINITY I TOTAL B-HCG REAGENT KIT SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN DHA ABBOTT IRELAND DIAGNOSTICS DIVISION 71475UD01 00380740163334

Patients

Seq Age Sex Outcome Treatment
1 21 YR Female ALNTY I PROCESSING MODU, 03R65-01,(B)(6)