IMP,TSV,3.7,10,MTX,MG
Report
- Report Number
- 0001038806-2025-03187
- Event Type
- Injury
- Date Received
- November 17, 2025
- Date of Event
- September 17, 2025
- Report Date
- January 15, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- UDI-DI
- 00889024019928
- PMA / PMN Number
- K101977
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A1: PATIENT IDENTIFIER: UNKNOWN / NOT PROVIDED. A4: PATIENT WEIGHT: UNKNOWN / NOT PROVIDED.
ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) TSVTB10, (IMP,TSV,3.7,10,MTX,MG) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE WAS SEEN ATTACHED TO THE EXTERNAL THREADS. THE IMPLANT WAS OBSERVED FRACTURED AT THE COLLAR REGION, AND A FRACTURED SCREW FRAGMENT WAS SEEN STUCK INSIDE THE IMPLANT. HOWEVER, CUSTOMER INDICATED THE SCREW FRACTURED DURING THE IMPLANT REMOVAL PROCESS. SOME DAMAGE WAS OBSERVED AROUND THE IMPLANT'S EXTERNAL THREADS LIKELY DUE TO THE REMOVAL PROCESS. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1240666. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1240666 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿ BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS, OR CUSTOMER ERROR. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
THE DOCTOR REPORTS THAT THE DENTAL IMPLANT LOCATED IN POSITION NUMBER 35 FAILED BECAUSE IT FRACTURED AT THE HEAD. THE DOCTOR REPORTS THAT THE PROCEDURE WAS CONCLUDED BY PLACING ANOTHER IMPLANT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362558 | IMP,TSV,3.7,10,MTX,MG | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 1240666 | 00889024019928 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Unknown | Required Intervention |