ALINITY C MAGNESIUM
Report
- Report Number
- 3005094123-2025-00575
- Event Type
- Malfunction
- Date Received
- November 17, 2025
- Date of Event
- November 11, 2025
- Report Date
- December 18, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- JGJ
- UDI-DI
- 00380740176532
- PMA / PMN Number
- K181748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM THE ALINITY C MAGNESIUM, 08P19-34, ABBOTT IRELAND DIAGNOSTICS DIVISION TO ALINITY C PROCESSING MODULE, 03R67-01, ABBOTT LABORATORIES. MDR NUMBER 3016438761-2025-00739 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C MAGNESIUM RESULT FOR AN 80 YEAR OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE: 1.6 TO 2.3 MG/DL): 11NOV2025, SID (B)(6): INITIAL RESULT = 8.3 MG/DL, REPEAT ON ANOTHER ALINITY = 2.0 MG/DL THE DISCREPANT RESULT WAS NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C MAGNESIUM RESULT FOR AN 80-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE: 1.6 TO 2.3 MG/DL): ON (B)(6) 2025, SID (B)(6): INITIAL RESULT = 8.3 MG/DL, REPEAT ON ANOTHER ALINITY = 2.0 MG/DL THE DISCREPANT RESULT WAS NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 228201 | ALINITY C MAGNESIUM | PHOTOMETRIC METHOD, MAGNESIUM | JGJ | ABBOTT IRELAND DIAGNOSTICS DIVISION | 74272UD01 | 00380740176532 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Male | ALNTY C PROCESSING MODU, 03R67-01, (B)(6).| ALNTY C PROCESSING MODU, 03R67-01, (B)(6). |