FDA Adverse Event Malfunction Summary report: N

ALINITY C MAGNESIUM

MDR report key: 23567619 · Received November 17, 2025

Report

Report Number
3005094123-2025-00575
Event Type
Malfunction
Date Received
November 17, 2025
Date of Event
November 11, 2025
Report Date
December 18, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
JGJ
UDI-DI
00380740176532
PMA / PMN Number
K181748
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM THE ALINITY C MAGNESIUM, 08P19-34, ABBOTT IRELAND DIAGNOSTICS DIVISION TO ALINITY C PROCESSING MODULE, 03R67-01, ABBOTT LABORATORIES. MDR NUMBER 3016438761-2025-00739 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C MAGNESIUM RESULT FOR AN 80 YEAR OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE: 1.6 TO 2.3 MG/DL): 11NOV2025, SID (B)(6): INITIAL RESULT = 8.3 MG/DL, REPEAT ON ANOTHER ALINITY = 2.0 MG/DL THE DISCREPANT RESULT WAS NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C MAGNESIUM RESULT FOR AN 80-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE: 1.6 TO 2.3 MG/DL): ON (B)(6) 2025, SID (B)(6): INITIAL RESULT = 8.3 MG/DL, REPEAT ON ANOTHER ALINITY = 2.0 MG/DL THE DISCREPANT RESULT WAS NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
228201 ALINITY C MAGNESIUM PHOTOMETRIC METHOD, MAGNESIUM JGJ ABBOTT IRELAND DIAGNOSTICS DIVISION 74272UD01 00380740176532

Patients

Seq Age Sex Outcome Treatment
1 80 YR Male ALNTY C PROCESSING MODU, 03R67-01, (B)(6).| ALNTY C PROCESSING MODU, 03R67-01, (B)(6).