FDA Adverse Event Malfunction Summary report: N

INSET II

MDR report key: 23566065 · Received November 17, 2025

Report

Report Number
3003442380-2025-16458
Event Type
Malfunction
Date Received
November 17, 2025
Date of Event
October 23, 2025
Report Date
November 11, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL (B)(4) - DEVICE 2 OF 2. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011953 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DHR REVIEW: THE LOT 6011953 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 121 AND MANUFACTURED IN THE LINE L4 LI39 ON 06-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING THE LOT 5B05638 WAS MANUFACTURED ACCORDING TO THE FORM VERSION 09 AND MANUFACTURED IN THE MACHINE ITL01, ON 07-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING THE LOT 5B00188 WAS MANUFACTURED ACCORDING TO THE FORM- VERSION 09 AND MANUFACTURED IN THE MACHINE ITL01, ON 02-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TWO INFUSION SET TUBING DETACHMENT EVENTS ON (B)(6) 2025. THE SITE OF DETACHMENT WAS FROM HUB. THE INFUSION SET WAS IN USE FOR COUPLE HOURS. THE BLOOD GLUCOSE LEVELS WAS HIGH AND PATIENT WAS TREATED WITH CORRECTION BOLUS VIA PUMP. THE PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274274 INSET II UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002817 6011953 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female