FDA Adverse Event Malfunction Summary report: N

TAPERLOC COMPLETE

MDR report key: 23564208 · Received November 15, 2025

Report

Report Number
0001825034-2025-03631
Event Type
Malfunction
Date Received
November 15, 2025
Date of Event
October 28, 2025
Report Date
March 26, 2026
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPZ
UDI-DI
00880304489776
PMA / PMN Number
K200196
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: D9; G3; H2; H3; H6; H10. THE FOLLOWING SECTIONS WERE CORRECTED: H4. VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAS NOT BEEN COMPROMISED. A DEVICE HISTORY REVIEW IS NOT REQUIRED. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. COMPLAINT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). G2: FOREIGN - EVENT OCCURRED IN JAPAN. THE CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING INSPECTION OF STOCK, THE STERILE PACKAGE WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2715436 TAPERLOC COMPLETE PROSTHESIS, HIP LPZ ZIMMER BIOMET, INC. 7459854 00880304489776

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown