FDA Adverse Event Malfunction Summary report: N

BECKMAN TOTAL BILIRUBIN TRIGGERED (TB-T) KIT

MDR report key: 2355639 · Received December 1, 2011

Report

Report Number
2050012-2011-08151
Event Type
Malfunction
Date Received
December 1, 2011
Date of Event
November 3, 2011
Report Date
November 3, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
CIG
PMA / PMN Number
K902801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TOTAL BILIRUBIN REAGENT IS CORROSIVE AND MAY CAUSE BURNS TO THE EYES, SKIN AND GASTROINTESTINAL TRACT AND SEVERE RESPIRATORY TRACT IRRITATION. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THERE WAS WATER IN THE REAGENT COMPARTMENT OF THE UNICEL DXC 800 SYNCHRON SYSTEM. CUSTOMER REPORTED THAT THERE THE LIQUID LOOKED CRYSTALLIZED WHERE IT HAD DRIED. CUSTOMER REPORTED THAT QUALITY CONTROL (QC) LOOKED GOOD WHEN THEY LAST RAN ABOUT AN HOUR AGO. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) FOUND THE TOTAL BILIRUBIN REAGENT CARTRIDGE WAS CRACKED AT THE BOTTOM. THE FSE PUT IN NEW REAGENT TO AVOID CONTAMINATION AND CALIBRATED THEN PERFORMED QUALITY CONTROL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN TOTAL BILIRUBIN TRIGGERED (TB-T) KIT DIAZO COLORIMETRY, BILIRUBIN CIG BECKMAN COULTER, INC.

Patients

Seq Age Sex Outcome Treatment
1