FDA Adverse Event Injury Summary report: N

EPIQ 7

MDR report key: 23547039 · Received November 13, 2025

Report

Report Number
3019216-2025-000433
Event Type
Injury
Date Received
November 13, 2025
Date of Event
October 23, 2025
Report Date
December 4, 2025
Manufacturer
PHILIPS ULTRASOUND, INC
Product Code
IYN
UDI-DI
00884838109995
PMA / PMN Number
K132304
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS WAS INITIALLY REPORTED FOR AN ALLEGATION OF A SERIOUS INJURY. HOWEVER, ADDITIONAL INFORMATION WAS RECEIVED THAT VERIFIED THERE WAS NO ACTUAL SERIOUS INJURY ASSOCIATED WITH THIS EVENT. THE ROUTINE TEE ECHOCARDIOGRAM WAS ABLE TO BE COMPLETED. IT WAS CLARIFIED THE PATIENT DID NOT FEEL ANY THROAT DISCOMFORT. THE EMOTIONAL TENSION CAUSED THE PATIENT'S BLOOD PRESSURE TO INCREASE TO AN UNSPECIFIED LEVEL BUT DID NOT REQUIRE ANY MEDICAL INTERVENTION OR AFFECT THE OUTCOME. THE REPORTED PROBLEM INDICATED IN THIS EVENT IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH IF IT WERE TO REOCCUR.

Description of Event or Problem · 0

PHILIPS RECEIVED AN NMPA#(B)(4). IT WAS REPORTED A CUSTOMER¿S EPIQ 7C ULTRASOUND SYSTEM, IN USE WITH AN X7-2T TRANSDUCER, WAS NOT AVAILABLE DUE TO A SYSTEM LOCK UP. THE SYSTEM WAS EXCHANGED, BUT THE TRANSDUCER COULD NOT BE USED AND REQUIRED A REPLACEMENT AS WELL. THE PATIENT EXPERIENCED THROAT DISCOMFORT AND EMOTIONAL EXCITEMENT. THE PATIENT'S BLOOD PRESSURE INCREASED TO AN UNSPECIFIED LEVEL. THE PATIENT REQUIRED REASSURANCE ALONG WITH INFORMATION REGARDING THE TRANSDUCER REPLACEMENT. THE CUSTOMER ALLEGES A SERIOUS INJURY OCCURRED, HOWEVER THERE IS INSUFFICIENT INFORMATION REGARDING THE PATIENT OUTCOME AT THIS TIME. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1929596 EPIQ 7 SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC IYN PHILIPS ULTRASOUND, INC EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM 00884838109995

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other