FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES

MDR report key: 23544979 · Received November 13, 2025

Report

Report Number
1917413-2025-01542
Event Type
Malfunction
Date Received
November 13, 2025
Date of Event
October 16, 2025
Report Date
October 23, 2025
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
GIM
UDI-DI
30382903678410
PMA / PMN Number
BK230980
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1: INITIAL REPORTER FACILITY NAME: (B)(6). THERE WERE MULTIPLE MEDICAL DEVICE TYPES: D2B: MEDICAL DEVICE TYPE: JKA THERE WERE MULTIPLE 510K NUMBERS G4: PMA / 510(K)#: K213670, K231373. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED ANY SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 100 RETAINED SAMPLES WERE VISUALLY INSPECTED, AND THE HEMOGARD CLOSURE ASSEMBLY WAS CORRECTLY ASSEMBLED AND PLACED ON THE TUBES. ADDITIONALLY, FUNCTIONAL TESTS WERE PERFORMED ON 10 RETAINED SAMPLES, AND ALL TUBES WERE WITHIN SPECIFICATION LIMITS WITH NO ISSUES IDENTIFIED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: UNDERFILL. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES, THE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2195772 BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE GIM BECTON, DICKINSON & CO. (BROKEN BOW) 4222893 30382903678410

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown