K-FILE M-ACCESS 25MM 010
Report
- Report Number
- 8031010-2025-00197
- Event Type
- Malfunction
- Date Received
- November 12, 2025
- Date of Event
- September 14, 2025
- Report Date
- November 25, 2025
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- EKS
- UDI-DI
- J003A12MA025010121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
WE RECEIVED THE FOLLOWING INFORMATION REGARDING THIS COMPLAINT: 1. THE PATIENT WAS NOT INJURED AND DID NOT REQUIRE TREATMENT. IT WAS THE DOCTOR'S EXCESSIVE FORCE DURING THE OPERATION THAT CAUSED THE INSTRUMENT TO SEPARATE. 2. THE PRODUCT HAS BEEN DISCARDED AND CANNOT BE RETURNED. INVESTIGATION RESULTS: SUMMARY: INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1905288). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE (HAND USED FILE - TECHNIQUE NO MORE VERIFIABLE TO DATE), OVERUSE (NUMBER OF USES NOT COMMUNICATED), EXCESSIVE WEAR, PATIENT CONDITION AND BEHAVIOR DURING TREATMENT OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE). IN ADDITION, THERE ARE FURTHER CIRCUMSTANCES ON THE CONDITIONS IN DENTISTRY (USED DISINFECTANTS, TRAINING/KNOWLEDGE STATUS), OR ANY OTHER ENVIRONMENTAL CONDITIONS, WHICH ARE UNKNOWN TO US AND MAY ALSO HAVE AN IMPACT ON THE REPORTED FAILURE MODE. THIS IS TO CORRECT AND REMOVE THE CODES THAT WERE INITIALLY REPORTED - REMOVING CODES FOR: HEALTH EFFECT - CLINICAL CODE - 4580, HEALTH EFFECT - IMPACT CODE - 4648. THE CORRECT CODES FOR THIS COMPLAINT ARE: HEALTH EFFECT - CLINICAL CODE - 4582, HEALTH EFFECT - IMPACT CODE - 2199. THIS IS A FOLLOW UP REPORT TO CORRECT THIS CODE.
AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. THE DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE LOT NUMBER WAS PROVIDED AND RETAINED-PRODUCT TESTING AND/OR DHR REVIEW ARE PLANNED. THE RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT, IT IS REPORTED THAT A K-FILE M-ACCESS 25MM 010 FILE BROKE DURING USE. THE BROKEN PART WAS RETRIEVED BY MICROSCOPE AND ULTRASONIC EQUIPMENT. CONTINUING WITH SUBSEQUENT TREATMENT. PATIENT INFORMATION PENDING. FURTHER INFORMATION REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2228445 | K-FILE M-ACCESS 25MM 010 | FILE, PULP CANAL, ENDODONTIC | EKS | MAILLEFER INSTRUMENTS HOLDING SARL | 1905288 | J003A12MA025010121 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |