BD PYXIS¿ ANESTHESIA STATION ES
Report
- Report Number
- 2016493-2025-133529
- Event Type
- Malfunction
- Date Received
- November 12, 2025
- Date of Event
- October 30, 2023
- Report Date
- November 5, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403533242
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING JANUARY 2022-DECEMBER 5, 2024, UNDER CAPA 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID LOCATE 1 SIMILAR COMPLAINT WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. ¿ CASE: (B)(4)¿ USER LOCKED OUT -AIW ANES1. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 19-APR-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE ACCOUNT WAS LOCKING UP AND PREVENTING USER LOGIN. A TECHNICAL SUPPORT SPECIALIST CONNECTED TO THE SERVER AND CONFIRMED THAT HOSPITAL IT MADE THE REQUIRED CHANGES TO THE AD USER. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS ANESTHESIA ES SYSTEM ACCOUNT WAS LOCKING UP AND THE USER COULDN'T LOG IN AND NURSES WERE UNABLE TO ISSUE THE MEDICATIONS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2545310 | BD PYXIS¿ ANESTHESIA STATION ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 327 | 10885403533242 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |