DAVINCI XI
Report
- Report Number
- 2955842-2025-44855
- Event Type
- Malfunction
- Date Received
- November 12, 2025
- Date of Event
- November 3, 2025
- Report Date
- November 19, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE IESU WAS ANALYZED AND THE REPORTED FAILURE (IESU FAULTS) WAS CONFIRMED USING ARTEMIS; C-34 ERRORS LOGGED ON (B)(6) 2025 WITH INITIAL OCCURRENCE AT 9:12 AM (ARTEMIS TIME). M-11, C-37, C-30 ERRORS LIKELY RELATED. ADDITIONAL M-1F, M-32, M-B1, M-35, M-36, M-1C ERRORS LOGGED IN ARTEMIS. INSPECTION DURING FA PROCESS WAS ABLE TO REPLICATE THE REPORTED EVENT; UNIT TESTED ON SYSTEM PRESENTS C-34/C-00 ERROR ON STARTUP ((B)(6) 2025 11:04 AM US-GA). ADDITIONAL C-00, M-0B, M-11 ERRORS IN ERBE LOGS. AS A RESULT OF THESE FINDINGS, THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED BECAUSE THE UNIT IS AN OEM PRODUCT AND CANNOT BE OPENED ON SITE FOR FURTHER ANALYSIS. ERBE WILL BE SENT TO OEM FOR FURTHER INVESTIGATION. THE COMPLAINT WAS CONFIRMED BASED ON THE FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO ELECTRICAL DEFECT OF THE IESU.
INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED GENERAL SURGERY SURGICAL PROCEDURE, THE CUSTOMER INFORMED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) THAT ERROR C-34 OCCURRED ON ERBE INTEGRATED ELECTROSURGICAL UNIT (IESU) WHEN THE SURGEON WAS ACTIVATING MONOPOLAR ENERGY. THE CUSTOMER POWER CYCLED THE IESU, BUT THE ISSUE WAS NOT RESOLVED. THE CUSTOMER ELECTED TO USE A THIRD-PARTY GENERATOR TO CONTINUE WITH THE PROCEDURE. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2592624 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-63 | N/A | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |