DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2011-00166
- Event Type
- Malfunction
- Date Received
- November 30, 2011
- Date of Event
- November 20, 2011
- Report Date
- November 20, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- DKJ
- PMA / PMN Number
- K904301
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED SALICYLATE RESULT WAS SAMPLE INTEGRITY. THE IFU FOR DIMENSION SALICYLATE FLEX REAGENT CARTRIDGES CONTAINS THE FOLLOWING INFORMATION: "FOLLOW THE INSTRUCTIONS WITH YOUR SPECIMEN COLLECTION DEVICE FOR USE AND PROCESSING." AND "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER." NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED SALICYLATE RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS RE-RUN ON AN ALTERNATE INSTRUMENT AND AN ELEVATED RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED SALICYLATE RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | SALICYLATE FLEX® REAGENT CARTRIDGE | DKJ | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | DA2183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |