FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDBANK TOWER

MDR report key: 23524678 · Received November 11, 2025

Report

Report Number
2016493-2025-133231
Event Type
Malfunction
Date Received
November 11, 2025
Date of Event
October 14, 2025
Report Date
October 17, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403517167
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 21-DEC-2018 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE USER WAS UNABLE TO SEE THE CHECKMARK WHEN TRYING TO PULL A CONTROLLED MEDICATION. THE TECHNICAL SUPPORT SPECIALIST (TSS) FOUND THAT A COWORKER HAD POSSIBLY REJECTED THE REQUEST, WHICH CAUSED IT TO BE BLOCKED FOR THE USER. THE REQUEST WAS RE-APPROVED AND VERIFIED UNDER RX VERIFY IN MY Q LINK CR. THE CUSTOMER CONFIRMED THAT EVERYTHING WAS WORKING CORRECTLY. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST RESOLVED THE ISSUE ON THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDBANK TOWER NO CHECKMARK APPEARED FOR NURSE WHEN TRYING TO PULL CONTROLLED MEDICATION. THE CUSTOMER STATED THAT THIS MALFUNCTION OCCURRED WHILE DISPENSING THE MEDICATION. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2645914 BD PYXIS¿ MEDBANK TOWER AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 500004500000 10885403517167

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown