PERCLOSE¿ PROSTYLE¿
Report
- Report Number
- 2024168-2025-11455
- Event Type
- Malfunction
- Date Received
- November 10, 2025
- Date of Event
- October 21, 2025
- Report Date
- December 16, 2025
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- UDI-DI
- 08717648344039
- PMA / PMN Number
- P960043
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED SUTURE DETACHMENT COULD NOT BE TESTED SOME AS THE SUTURE WAS NOT RETURNED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, A DEFINITIVE CAUSE FOR THE REPORTED ISSUE COULD NOT BE DETERMINED. FACTORS THAT MAY CONTRIBUTE TO SUTURE SEPARATION DURING KNOT ADVANCEMENT INCLUDE, BUT ARE NOT LIMITED TO, USER ERROR (PUSHING MORE THAN TENSIONING THE RAIL LIMB, LEVER DEPLOYED EARLY OR EXCESSIVE SUTURE TENSION) OR SUTURE/ NICK DAMAGE. THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. CORRECTIONS: D4 - LOT #: UPDATED FROM 4091041 TO 5051041. D4 - EXPIRATION DATE: UPDATED FROM 8/31/2026 TO 4/30/2027. D4 - PRIMARY UDI NUMBER: UPDATED FROM (B)(4) TO (B)(4). H4 - DEVICE MFR DATE: UPDATED FROM 9/10/2024 TO 5/10/2025.
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE CALCIFIED RIGHT COMMON FEMORAL ARTERY USING A PROSTYLE DEVICE AFTER A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE USING A 6F SHEATH. REPORTEDLY, A SUTURE BREAK OCCURRED WHEN PULLING THE BLUE RAIL SUTURE AND USING THE SUTURE TRIMMER TO ADVANCE THE KNOT . MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2615404 | PERCLOSE¿ PROSTYLE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12773-02 | 5051041 | 08717648344039 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |