FDA Adverse Event Malfunction Summary report: N

CAVITRON SELECT SPS G124 GENERATOR(DNA)

MDR report key: 23512397 · Received November 10, 2025

Report

Report Number
2424472-2025-00156
Event Type
Malfunction
Date Received
November 10, 2025
Report Date
December 8, 2025
Manufacturer
DENTSPLY LLC
Product Code
ELC
UDI-DI
D003813021
PMA / PMN Number
K970123
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(6) 2025 EVALUATION TECH: (B)(6). W4D WATER SOL, ISO VALVE BUILD UP/DEBRIS DEBRIS BUILDUP IN THE WATER SOLENOID, HARDENED AND DETERIORATED WATER SUPPLY HOSE, DEBRIS BUILDUP IN THE WATER FILTER. BLOCKAGE IN THE HANDPIECE CABLE CAUSING RESTRICTED WATER FLOW. WILL REPLACE DAMAGED/WORN COMPONENTS AND RECALIBRATE UNIT TO FACTORY SPECS UPON ESTIMATE APPROVAL. NO HANDPIECE RECEIVED FOR EVALUATION. HPC: 0423. HP: N/A. DHR REVIEW IS NOT REQUIRED BECAUSE THE PRODUCT WAS RETURNED FOR EVALUATION, AND THE DESCRIBED FAILURE MODE IS A KNOWN HAZARD.

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE LACK OF WATER FLOW HAS CAUSED AN OVERHEATING INSERT. SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803.

Description of Event or Problem · 0

WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1928294 CAVITRON SELECT SPS G124 GENERATOR(DNA) SCALER, ULTRASONIC ELC DENTSPLY LLC D003813021

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown