SMR SHOULDER
Report
- Report Number
- 3008021110-2025-00142
- Event Type
- Injury
- Date Received
- November 10, 2025
- Date of Event
- September 16, 2025
- Report Date
- November 10, 2025
- Manufacturer
- LIMACORPORATE S.P.A.
- Product Code
- PHX
- UDI-DI
- 08033390204538
- PMA / PMN Number
- K212800
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
CHECKING THE MANUFACTURING CHARTS OF THE COMPONENTS REMOVED, NO PRE-EXISTING ANOMALIES HAVE BEEN FOUND IN THE ITEMS BELONGING TO THE SAME PRODUCT CODE AND LOT NUMBER AS THE DEVICES INVOLVED IN THE REVISION SURGERY HEREBY REPORTED. THE MANUFACTURER WILL SUBMIT A FINAL MDR AS SOON AS THE INVESTIGATION IS COMPLETE.
REVISION SURGERY DUE TO INSTABILITY PERFORMED ON (B)(6) 2025. PATIENT UNDERWENT SURGERY FOR A REVERSE SHOULDER ON (B)(6) 2025, IN WHICH THE FOLLOWING COMPONENTS WERE IMPLANTED: - SHORT STEM PLUS #5 (PART CODE 1357.14.236, LOT NUMBER 2413552, STERILIZATION (B)(4)) - PRIMA REVERSE TRAY #SHORT (PART CODE 1367.15.010, LOT NUMBER 2434587, STERILIZATION (B)(4)) - PRIMA REV.INSERT #SHORT D.42MM (PART CODE 1367.54.300, LOT NUMBER 2122627, STERILIZATION (B)(4)) - SMR GLENOSPHERE Ø 42MM (PART CODE 1374.09.131, LOT NUMBER 2123235, STERILIZATION (B)(4)) - SMR SMALL-R CONNECTOR +4 (PART CODE 1374.15.314, LOT NUMBER 2101336, STERILIZATION (B)(4)) - SMR METAL-BACK GLENOID SMALL R (PART CODE 1375.21.005, LOT NUMBER 2402374, STERILIZATION (B)(4)) THUMB TEST WAS DONE ON HUMERAL CUT SURFACE PRIOR TO IMPLANTATION OF PRIMA STEM. DUE TO THE SIZE OF THE PATIENT, THE SURGEON DETERMINED THE NECESSITY OF THIS IMPLANT TO ACHIEVE DESIRED MEDIALIZATION OF THE HUMERUS AND LATERALIZATION OF THE GLENOID COMPONENTS. THEREFORE, DURING THE REVISION HEREBY REPORTED, THE GLENOSPHERE AND THE LATERALIZED CONNECTOR WERE REMOVED AND REPLACED BY THE FOLLOWING COMPONENTS: - SMR GLENOSPHERE Ø 40MM (PART CODE 1374.09.121, LOT NUMBER 2400301, STERILIZATION (B)(4)) - SMR SMALL-R CONNECTOR +2 (PART CODE 1374.15.312, LOT NUMBER 2410294, STERILIZATION (B)(4)) THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1963. THE EVENT HAPPENED IN THE UNITED STATES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2581295 | SMR SHOULDER | SMR SHOULDER - GLENOSPHERE DIA. 42 MM | PHX | LIMACORPORATE S.P.A. | 1374.09.131 | 2123235 | 08033390204538 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |