FDA Adverse Event Injury Summary report: N

SHOULDER SYSTEM

MDR report key: 23510653 · Received November 10, 2025

Report

Report Number
3005180920-2025-01127
Event Type
Injury
Date Received
November 10, 2025
Date of Event
October 24, 2025
Report Date
November 10, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706285
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 28 OCTOBER 2025 REVERSE SHOULDER SYSTEM 04.01.0121 HUMERAL REVERSE HC LINER D 36/+6MM (K170452) LOT. 2501309: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17 APRIL 2025. EXPIRATION DATE: 03 APRIL 2030. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 27 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0278 SENSITIN LAT. GLENOSPHERE 36XD 24.5 (K203755) LOT. 2012762: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18 MARCH 2021. EXPIRATION DATE: 08 MARCH 2026. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 32 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH A SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE NO DEFINITIVE ROOT CAUSE CAN BE ESTABLISHED, WHILE THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

THE PATIENT INITIALLY UNDERWENT PRIMARY SHOULDER SURGERY USING A COMPETITOR & RSQUO, S PRODUCTS. ON (B)(6) 2025, A REVISION WAS PERFORMED WITH MEDACTA IMPLANTS. ON (B)(6), 2025, THE PATIENT PRESENTED WITH PAIN DUE TO A DISLOCATION OF UNKNOWN CAUSE. REVISION SURGERY WAS PERFORMED TO ENLARGE THE GLENOSPHERE, METAPHYSIS, AND LINER TO ENHANCE JOINT STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2615170 SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø36/+6MM PHX MEDACTA INTERNATIONAL SA 04.01.0121 2501309 07630040706285

Patients

Seq Age Sex Outcome Treatment
1 59 YR Female Required Intervention