FDA Adverse Event Malfunction Summary report: N

BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM

MDR report key: 23507708 · Received November 10, 2025

Report

Report Number
2016493-2025-131845
Event Type
Malfunction
Date Received
November 10, 2025
Date of Event
July 26, 2022
Report Date
October 30, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512568
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING JANUARY 2022-DECEMBER 5, 2024, UNDER CAPA: 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN: (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN: (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 08-MAY-2019 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE KEYBOARD NUMBERS WERE NOT WORKING. A FIELD SERVICE ENGINEER INSTALLED A NEW USB KEYBOARD. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM KEYBOARD NUMBERS WERE NOT WORKING WHEN I TRIED TO SIGN IN OR ISSUE MEDICATIONS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171814 BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 169-40 10885403512568

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown