STEALTHSTATION¿ S8 SYSTEM
Report
- Report Number
- 1723170-2025-03679
- Event Type
- Malfunction
- Date Received
- November 7, 2025
- Date of Event
- October 17, 2025
- Report Date
- November 18, 2025
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- UDI-DI
- 00763000272739
- PMA / PMN Number
- K162309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9 735821R, SERIAL/LOT #: -, UBD: , UDI#: H3, H6; THE CAMERA, LOT NUMBER: P902977, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED PSU HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY DATING BACK TO DEC 2019, AND INTERMITTENT ILLUMINATOR CURRENT LOW DATING BACK TO FEB 2020. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .576MM WITH A PASSING THRESHOLD OF .250MM. THIS PSU HAD NOT BEEN PREVIOUSLY RETURNED FOR A FAILURE. CODES B01, C02, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. HARDWARE WAS REPLACED. CODES B01, C02, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHILE SETTING UP FOR A CASE, THE SITE'S INSTRUMENTS WERE NOT BEING TRACKED WHEN HELD IN FRONT OF THE CAMERA AND THE SYSTEM WAS DISPLAYING LOCALIZER NOT CONNECTED. THE SITE SWITCHED TO ANOTHER SYSTEM, AND THE SAME ISSUE OCCURRED. THERE WAS NO PATIENT INVOLVEMENT. ON 2025-OCT-17 IUD (REP): ADDITIONAL INFORMATION WAS RECEIVED. A MANUFACTURER REPRESENTATIVE (REP) VISITED THE SITE AND FOUND THAT THE MESSAGE DISPLAYED ON THIS SYSTEM WAS LOCALIZER FAULTED. TECHNICAL SERVICES (TS) HAD THE REP CHECK NDITOOLBOX. THE FOLLOWING FINDINGS WERE OBSERVED: BUMP DETECTOR BATTERY FAULT, BUMP DETECTED, AND STORAGE TEMPERATURE EXCEEDED. THE INTERNAL POSITIONING SENSOR UNIT (PSU) BATTERY HAS LIKELY FAILED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1873734 | STEALTHSTATION¿ S8 SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. | 9735665 | 00763000272739 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |