TANGILE FILL
Report
- Report Number
- 1417519-2025-00005
- Event Type
- Injury
- Date Received
- November 6, 2025
- Date of Event
- July 31, 2025
- Report Date
- October 23, 2025
- Manufacturer
- AMSINO HEALTHCARE (USA) INC.
- Product Code
- MRC
- UDI-DI
- 00198168143803
- PMA / PMN Number
- K201069
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE END USER RETURNED (B)(4) OF THE (B)(4) VIALS FROM THE CASE, WHICH AMSINO RECEIVED ON 14-AUG-2025. INITIAL FINISHED-PRODUCT TESTING FOR LOT L095, INCLUDED ASSAY, USP <71> STERILITY, AND USP <85> BACTERIAL ENDOTOXINS TESTING. TO OBTAIN MULTI-LABORATORY CONFIRMATION, AMSINO SUBMITTED A SUBSET OF THE RETURNED COMPLAINT SAMPLES FROM L095 TO A DIFFERENT LABORATORY FOR USP <71> STERILITY AND USP <85> ENDOTOXIN TESTING. USP <85> RESULTS WERE RECEIVED ON 22-OCT-2025 AND MET SPECIFICATION; USP <71> RESULTS WERE RECEIVED ON 31-OCT-2025 AND WERE NEGATIVE. IN ADDITION, RETENTION SAMPLES FROM THE SAME LOT WERE TESTED USING THE SAME METHODS, AND ALL RESULTS WERE WITHIN SPECIFICATION.
THE END USER REPORTED: I WAS A USER OF NUTRIFILL UNTIL I GOT THE RECALL NOTICE. I DISPOSED OF ALL REMAINING NUTRIFILL PRODUCT AS RECOMMENDED. UNFORTUNATELY, I NEEDED TO REPLACE MY SCLERAL SOLUTION RIGHT AWAY AND IT WOULD HAVE TAKEN TOO LONG FOR IT TO ARRIVE FROM THE TANGIBLE SCIENCE WEBSITE. I ORDERED IT FROM AMAZON SO THAT I COULD GET IT ASAP SO I COULD CONTINUE WEARING MY SCLERALS (REQUIRED FOR MY VISION). THE FIRST DAY I USED TANGIBLE FILL, MY EYES WERE SLIGHTLY IRRITATED BY THE END OF THE DAY. I ASSUMED THIS WAS JUST ADJUSTING TO THE NEW PRODUCT. BY DAY 5, MY EYES WERE INFECTED (RED, DRAINING, AND GUNKY). FOLLOW UP MESSAGE FROM END USER: I GOT SOME EYE MEDICATION AND THE INFECTION HAS CLEARED UP. I STARTED USING A DIFFERENT PRODUCT (BRAND) AND HAVE HAD NO FURTHER ISSUES. I HAVE THE REMAINING PRODUCT IF YOU NEED TO HAVE IT RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173832 | TANGILE FILL | SCLERAL SALINE 5ML | MRC | AMSINO HEALTHCARE (USA) INC. | 2490 | L095 | 00198168143803 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other| R |