FDA Adverse Event Malfunction Summary report: N

VOALTE NURSE CALL

MDR report key: 23490698 · Received November 6, 2025

Report

Report Number
2027454-2025-00202
Event Type
Malfunction
Date Received
November 6, 2025
Date of Event
October 10, 2025
Report Date
January 6, 2026
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
ILQ
UDI-DI
00887761995079
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING, HOWEVER IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION, THE ADDITIONAL RELEVANT INFORMATION WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

THE BAXTER TECHNICIAN FOUND IMPROPER CONFIGURATION CHANGES ON SECONDARY MONITORING SETTINGS WHICH WERE MADE BY THE SITE. THE VOALTE NURSE CALL SYSTEM PROVIDES A COMPREHENSIVE COMMUNICATION AND INFORMATION SYSTEM THAT PLACES PATIENT CALLS, STAFF CALLS AS WELL AS EMERGENCY AND CODE CALLS. THE SYSTEM PROVIDES AN ANNUNCIATION OF THESE CALLS AT BOTH ROOM LOCATIONS AND PRIMARY(DEDICATED) NURSE CALL STATIONS. THE VOALTE NURSE CALL SYSTEM ALSO HAS AN OPTION FOR SECONDARY NOTIFICATION CALLS TO CUSTOMERS PROVIDED THIRD-PARTY PRODUCTS (E.G CELL PHONES) WHERE CALLS MAY ALSO BE FORWARDED. THE SECONDARY NOTIFICATION WORKFLOWS ARE OPTIONS OFFERED TO THE FACILITIES THAT THE FACILITY MAY OR MAY NOT CHOOSE TO IMPLEMENT AS AN ADD ON TO THEIR PRIMARY NOTIFICATIONS, DEPENDING ON THEIR FACILITY¿S DESIGN AND NEEDS. THE NURSE CALL SYSTEM SECONDARY NOTIFICATION PROVIDES VISUAL AND/OR AUDIBLE EVENT ALERT NOTIFICATION VIA CUSTOMER DESIGNATED NURSE CALL COMMUNICATION DEVICES. THE LOCATION OF THESE DEVICES CAN BE AT A SPECIFIC CONSOLE IN PBX DEPARTMENT, A STAFF STATION LOCATED IN A BREAK ROOM OR HALLWAY, MOBILE PHONES, ETC. THE CUSTOMER PREFERENCE/REQUEST AT THE TIME OF THE INITIAL INTEGRATION. THE BAXTER TECHNICIAN ARRANGED TO GET THE ROOM CONFIGURED CORRECTLY FOR THE CUSTOMER TO RESOLVE THE ISSUE. ALTHOUGH THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT, IF THE REPORTED EVENT WERE TO RECUR IT COULD LEAD TO INJURY OR DEATH. THEREFORE, BAXTER IS REPORTING THIS COMPLAINT.

Description of Event or Problem · 0

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE STAFF EMERGENCY TRIGGERED FROM THE ROOM BUT NOT ALARMED. THERE WAS NO PATIENT/USER INJURY REPORTED.

Description of Event or Problem · 0

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE STAFF EMERGENCY TRIGGERED FROM THE ROOM BUT NOT ALARMED. THERE WAS NO PATIENT/USER INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176211 VOALTE NURSE CALL SYSTEM, COMMUNICATION, POWERED ILQ BAXTER HEALTHCARE CORPORATION P2599 00887761995079

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown N/A.