EVERA MRI XT DR SURESCAN
Report
- Report Number
- 9614453-2025-04461
- Event Type
- Injury
- Date Received
- November 6, 2025
- Date of Event
- October 14, 2025
- Report Date
- November 6, 2025
- Manufacturer
- MEDTRONIC EUROPE SARL
- Product Code
- LWS
- UDI-DI
- 00763000059521
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CONTINUATION OF D10: 0180 LEAD IMPLANTED: (B)(6) 2007, 407645 LEAD IMPLANTED (B)(6) 2005. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH EROSION. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1562683 | EVERA MRI XT DR SURESCAN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) | LWS | MEDTRONIC EUROPE SARL | DDMB1D1 | 00763000059521 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Female | Required Intervention| H |