FDA Adverse Event Injury Summary report: N

NICVIEW 2.0

MDR report key: 23485726 · Received November 6, 2025

Report

Report Number
9612197-2025-00008
Event Type
Injury
Date Received
November 6, 2025
Date of Event
October 28, 2025
Report Date
December 10, 2025
Manufacturer
NATUS MEDICAL INCORPORATED
Product Code
FWC
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORT REF NATUS COMPLAINT # (B)(4). REPLACEMENT NVARM2 SHIPPED TP CUSTOMER (B)(6) 2025 (B)(4). DEFECTIVE PART REQUESTED FOR RETURN (B)(6). DEFECTIVE PART REQUESTED FOR RETURN (B)(6). DEFECTIVE PART REQUESTED FOR RETURN (B)(6). THE CUSTOMER NEVER RETURNED THE DEFECTIVE PART BACK; NO INVESTIGATION COULD BE CARRIED OUT. RISK MANAGEMENT FILE REVIEW: DOC-023354 REV B - RISK ANALYSIS FOR NICVIEW 2.0 HAZARD 6.7. CAUSE - ARM ASSEMBLY FAILS DUE TO EXCESSIVE STATIC LOADS OR WEAR AND SAGS. EFFECT - SYSTEM FALL. RESIDUAL RISK - MODERATE. SEVERITY - 11. THE HAZARDS IDENTIFIED HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH A KNOWN STANDARD. AS NOTED IN QMS-000018 RISK MANAGEMENT SYSTEM PROCEDURE (SENSORY), HAZARDS THAT HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH KNOWN STANDARDS CAN BE PRESUMED TO BE CONSISTENT WITH AN ACCEPTABLE LEVEL OF RISK, YIELDING AN ACCEPTABLE RISK / BENEFIT TO THE PATIENT OR USER. RISK OUTWEIGHED BY BENEFIT OF USE OF DEVICE.

Additional Manufacturer Narrative · 0

INITIAL REPORT REF NATUS COMPLAINT # (B)(4). REPLACEMENT NVARM 2 SHIPPED TO CUSTOMER 4 NOV 2025 ((B)(4)) DEFECTIVE PART REQUESTED FOR RETURN.

Description of Event or Problem · 0

THE CAMERA WASN'T FUNCTIONING, SO THE EMPLOYEE DISCONNECTED AND CONNECTED/ PLUGGED IT BACK IN. AFTER, THE CAMERA SWUNG DOWN AND HIT HER ON THE BACK.

Description of Event or Problem · 0

THE CAMERA WASN'T FUNCTIONING, SO THE EMPLOYEE DISCONNECTED AND CONNECTED / PLUGGED IT BACK IN. AFTER, THE CAMERA SWUNG DOWN AND HIT HER ON THE BACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2552146 NICVIEW 2.0 NICVIEW 2.0 FWC NATUS MEDICAL INCORPORATED NVARM2

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown