FDA Adverse Event Malfunction Summary report: N

THERMOGARD HQTM CONSOLE

MDR report key: 23480674 · Received November 5, 2025

Report

Report Number
3010617000-2025-00759
Event Type
Malfunction
Date Received
November 5, 2025
Date of Event
October 9, 2025
Report Date
November 5, 2025
Manufacturer
ZOLL CIRCULATION, INC.
Product Code
NCX
UDI-DI
00849111075800
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FOR G4: PMA/510(K): PREMARKET SUBMISSION NUMBER NOT AVAILABLE/NOT RELEASED. THE CUSTOMER'S COMPLAINT THAT THE START-UP KIT (SUK) LFL TUBING AROUND THE ROLLER PUMP RUPTURED, CAUSING SALINE TO SPURT FROM THE TUBING INTO THE ROLLER PUMP OF THE THERMOGARD HQ CONSOLE (SN (B)(6) WAS CONFIRMED DURING THE VISUAL INSPECTION. THE ROLLER PUMP KNOB ON THE CONSOLE WAS THE OLD TYPE, WHICH CAUSED THE SUK TUBE TO RUPTURE. UPON VISUAL INSPECTION, IT WAS CONFIRMED THAT NORMAL SALINE HAD ACCUMULATED IN THE PUMP RACEWAY. THE CONSOLE INTERIOR WAS ALSO INSPECTED, BUT NO INFILTRATION OF NORMAL SALINE WAS OBSERVED. THE SALINE TRACES WERE CLEANED, AND THE OLD ROLLER PUMP KNOB WAS REPLACED WITH THE NEW VERSION TO ADDRESS THE REPORTED COMPLAINT. FOLLOWING CLEANING AND PART REPLACEMENT, THE THERMOGARD HQ CONSOLE PASSED THE FUNCTIONAL TESTING WITHOUT ERROR. FOLLOWING SERVICE, THE THERMOGARD HQ CONSOLE PASSED THE FINAL FUNCTIONAL AND ELECTRICAL SAFETY TESTS WITHOUT ANY ISSUES. HISTORICAL COMPLAINTS WERE REVIEWED FOR SERVICE INFORMATION RELATED TO THE REPORTED COMPLAINT, AND THERE WAS NO PREVIOUS HISTORY OF COMPLAINTS REPORTED FOR THE THERMOGARD HQ CONSOLE WITH SN (B)(6).

Description of Event or Problem · 0

DURING IVTM THERAPY, THE START-UP KIT (LOT # 200280) LFL TUBING AROUND THE ROLLER PUMP RUPTURED, CAUSING SALINE TO SPURT FROM THE TUBING INTO THE ROLLER PUMP OF THE THERMOGARD HQ CONSOLE (SN (B)(6). BECAUSE THE ROLLER PUMP WAS FILLED WITH SALINE, THE CUSTOMER STOPPED USING THE THERMOGARD HQ CONSOLE AND SWITCHED TO A SYSTEM FROM ANOTHER MANUFACTURER TO CONTINUE TEMPERATURE MANAGEMENT. THE ROLLER PUMP KNOB ON THE CONSOLE WAS OF THE OLD TYPE. NO PATIENT INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2491799 THERMOGARD HQTM CONSOLE INTRAVASCULAR CORE TEMPERATURE REGULATION SYSTEM NCX ZOLL CIRCULATION, INC. THERMOGARDHQ, LANG SET A 00849111075800

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown