FDA Adverse Event Death Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 23474844 · Received November 5, 2025

Report

Report Number
3008642652-2025-10708
Event Type
Death
Date Received
November 5, 2025
Date of Event
October 17, 2025
Report Date
November 4, 2025
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION OF THE MONITOR AND ELECTRODE BELT HAS BEEN COMPLETED. THE EVALUATION INCLUDED REVIEW OF DOWNLOADED SOFTWARE FLAG FILES ON THE DAY OF THE EVENT AND INCOMING FUNCTIONAL TESTING. DURING THE INCOMING FUNCTIONAL TESTING, A 1HZ SIMULATED NORMAL SINUS RHYTHM SIGNAL WAS APPLIED TO THE ECG ELECTRODES, FOLLOWED BY A 5HZ SIMULATED TREATABLE ARRHYTHMIA SIGNAL WHICH VERIFIED PROPER PERFORMANCE OF THE DETECTION ALGORITHM. DURING THE TRANSITION TO THE 5HZ SIGNAL, THE DEVICE WAS CONFIRMED TO PROPERLY ENTER INTO A TREATMENT SEQUENCE WHICH INCLUDES A VERIFICATION OF THE TACTILE VIBRATION ALARM, AUDIO MESSAGING, AND SIREN ALARMS, AS WELL AS A TEST OF THE PULSE DELIVERY CIRCUITRY. THE PULSE DELIVERY CIRCUITRY TEST VERIFIED PROPER CHARGING OF THE HIGH VOLTAGE CAPACITORS AND PROPER DELIVERY OF FIVE FULL ENERGY 150J BIPHASIC PULSES. THE FUNCTIONAL TESTING CONFIRMED PROPER RESPONSE BUTTON FUNCTIONALITY, ECG ACQUISITION, DETECTION ALGORITHM PERFORMANCE, AND PULSE DELIVERY FUNCTIONALITY. THERE IS NO INDICATION OF A PRODUCT MALFUNCTION.

Description of Event or Problem · 0

A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT A PATIENT PASSED AWAY ON (B)(6) 2025 WHILE REPORTEDLY WEARING THE LIFEVEST. THE PATIENT RECEIVED AN INAPPROPRIATE TREATMENT. THE DEVICE WAS STARTED UP AT 18:18:58 ON (B)(6) 2025. AT 10:25:37 ON (B)(6) 2025, AN ARRHYTHMIA WAS DETECTED. ECG SHOWS AF @ 80 BPM WITH PVC¿S AND TALL T WAVES. AT 10:26:50, THE PATIENT RECEIVED THE INAPPROPRIATE TREATMENT. DOUBLE COUNTING CONTRIBUTED TO THE FALSE DETECTION. RHYTHM AT THE TIME OF TREATMENT WAS AF @ 80 BPM WITH PVC¿S AND TALL T WAVES. POST SHOCK RHYTHM WAS AF @ 80 BPM WITH PVC¿S. THE ELECTRODE BELT WAS DISCONNECTED AT 10:29:26 ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2491372 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM 00855778005005

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Death