FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 23471242 · Received November 4, 2025

Report

Report Number
2955842-2025-43840
Event Type
Malfunction
Date Received
November 4, 2025
Date of Event
October 1, 2025
Report Date
December 23, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874119792
PMA / PMN Number
K214095
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A FRAYED GRIP CABLE. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE. THE PROBABLE ROOT CAUSE OF A FRAYED CABLE IS ATTRIBUTED TO DAMAGE TO THE CABLE THROUGH EXTERNAL COLLISIONS, DURING USE, OR DURING REPROCESSING. THIS ISSUE CAN BE RESOLVED BY USING AN ALTERNATE INSTRUMENT TO COMPLETE THE PROCEDURE.

Additional Manufacturer Narrative · 0

THE INSTRUMENT INVOLVED IN THIS COMPLAINT HAS NOT BEEN RECEIVED AND/OR TESTED BY FAILURE ANALYSIS AS OF THE DATE OF THIS REPORT. IF THE PRODUCT IS RECEIVED AND EVALUATED, A SUPPLEMENTAL MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSTRUMENT STOPPED WORKING PROPERLY, SO IT WAS REMOVED FOR EXAMINATION. IT WAS OBSERVED THAT ONE OF THE PULLEYS CABLES WAS CUT, SO IT WAS REPLACED WITH A BACKUP INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1745082 ENDOWRIST MARYLAND BIPOLAR FORCEPS NAY INTUITIVE SURGICAL, INC 471172-19 K10250327 0362 00886874119792

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES