FDA Adverse Event Injury Summary report: N

COMP AUG MINI BSPLT W TPR MD

MDR report key: 23468660 · Received November 4, 2025

Report

Report Number
0001825034-2025-03460
Event Type
Injury
Date Received
November 4, 2025
Date of Event
May 23, 2025
Report Date
November 4, 2025
Manufacturer
ZIMMER BIOMET, INC.
Product Code
PHX
UDI-DI
00880304993242
PMA / PMN Number
K172502
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). D10: MEDICAL PRODUCTS: ITEM#: 110032420, COMP AUG MINI BSPLT W TPR MD; LOT#: 64155087. ITEM#: 115316, COMP RVRS SHLDR GLNSP +6 36MM; LOT#: 943630. G2: FOREIGN: AUSTRALIA. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: A REVERSE-TYPE RIGHT SHOULDER ARTHROPLASTY IS PRESENT. THE GLENOSPHERE IS SEPARATED FROM THE GLENOID BASEPLATE. THERE IS A RESULTING DISLOCATION OF THE IMPLANTS. THE HUMERAL STEM IS INCOMPLETELY VISUALIZED. NO FRACTURE IS IDENTIFIED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL SHOULDER ARTHROPLASTY APPROXIMATELY FIVE (5) YEARS AND FIVE (5) MONTHS AGO. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION SURGERY APPROXIMATELY ONE (1) MONTH AGO DUE TO THE IMPLANTS DISASSOCIATING. IT IS UNKNOWN AT THIS TIME WHAT CAUSED THE IMPLANTS TO DISASSOCIATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1744488 COMP AUG MINI BSPLT W TPR MD SHOULDER PROSTHESIS, REVERSE CONFIGURATION PHX ZIMMER BIOMET, INC. 66747368 00880304993242

Patients

Seq Age Sex Outcome Treatment
1 73 YR Unknown Hospitalization| R SEE H11 NARRATIVE.