FDA Adverse Event Malfunction Summary report: N

BIO CONSOLE 560

MDR report key: 23467212 · Received November 4, 2025

Report

Report Number
2184009-2025-01416
Event Type
Malfunction
Date Received
November 4, 2025
Date of Event
October 29, 2025
Report Date
January 7, 2026
Manufacturer
PERFUSION SYSTEMS
Product Code
DWA
UDI-DI
00763000236526
PMA / PMN Number
K080824
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE BATTERY WAS INSTALLED IN THE INSTRUMENT IN 2022 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

B.5.MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT BATTERY FAILURE WAS VERIFIED THROUGH TESTING.THE LOT NUMBERS FOUND ON THE ORIGINAL BATTERIES WAS 2223/ 190222-M MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: THE REPORTED BATTERY FAILURE ISSUE WAS VERIFIED DURING SERVICE. THE ISSUE WAS RESOLVED BY REPLACING THE B ATTERIES. PREVENTIVE MAINTENANCE WAS PERFORMED AS PER SPECIFICATION. D9: INSTRUMENT WAS SERVICED AT THE FACILITY BY MEDTRONIC FIELD SERVICE AND WAS NOT RETURNED TO MEDTRONIC SERVICE CENTER. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-CONSOLE 560 CONTROLLER INSTRUMENT, IT WAS REPORTED THAT THE BATTERY FAILED. THE INSTRUMENT WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1698740 BIO CONSOLE 560 CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS DWA PERFUSION SYSTEMS 560BCS1 00763000236526

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown