FDA Adverse Event Malfunction Summary report: N

BECKMAN TOTAL BILIRUBIN TRIGGERED KIT

MDR report key: 2346698 · Received November 23, 2011

Report

Report Number
2050012-2011-07858
Event Type
Malfunction
Date Received
November 23, 2011
Date of Event
October 26, 2011
Report Date
November 9, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
CKA
PMA / PMN Number
K883181
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER INC. ASSESSMENT OF THE SUPPLIED PHOTOS INDICATED THAT THE CARTRIDGES WERE LEAKING FROM THE BOTTOM SEAM. THIS MAY BE INDICATIVE OF A CARTRIDGE MANUFACTURING ISSUE. NO SERVICE WAS DISPATCHED TO THE SITE FOR THIS EVENT. A DEFINITIVE ROOT CAUSE FOR THE LEAK IS NOT KNOWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY OBSERVED A LEAK COMING FROM A SYNCHRON CX4/CX5 CLINICAL CHEMISTRY ANALYZER REAGENT. NO PATIENT RESULTS WERE INVOLVED IN THIS EVENT. IT IS UNKNOWN AS TO WHETHER THE PERSONNEL INVOLVED IN THIS EVENT WAS WEARING PERSONAL PROTECTIVE EQUIPMENT AT THE TIME OF OCCURRENCE HOWEVER NO PERSONNEL SOUGHT MEDICAL ATTENTION IN ASSOCIATION WITH THIS EVENT. THERE WAS NO CHEMICAL EXPOSURE TO PERSONNEL UNCOVERED WOUNDS OR MUCOUS MEMBRANES. NO DEATH, INJURY OR MODIFICATION TO PATIENT TREATMENT WAS ASSOCIATED WITH THIS EVENT. NO EXPOSURE CONTROL PLAN WAS IN PLACE AT THE FACILITY. THE MATERIAL SAFETY DATA SHEET WAS NOT REVIEWED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN TOTAL BILIRUBIN TRIGGERED KIT NADH OXIDATION/NAD REDUCTION, ALT/SGPT CKA BECKMAN COULTER, INC. NA Z104209