FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ MYCO/F LYTIC CULTURE VIALS

MDR report key: 23465559 · Received November 4, 2025

Report

Report Number
2647876-2025-00658
Event Type
Malfunction
Date Received
November 4, 2025
Date of Event
October 15, 2025
Report Date
January 13, 2026
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904422882
PMA / PMN Number
K970333
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CATALOG 442288. BATCH NO. 5057481. CUSTOMER REPORTED A FALSE NEGATIVE RESULT. PHOTO OF A BOTTLE WITHOUT SENSOR WAS RECEIVED. BD WAS UNABLE TO REPRODUCE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. SATISFACTORY RESULTS WERE OBTAINED FROM RETENTION SAMPLES WHEN TESTED FOR MICROBIAL INSTRUMENT DETECTION AS PER QUALITY PROCEDURES. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT (I.E. FALSE NEGATIVE). MICROBIAL INSTRUMENT DETECTION INDICATED THAT THE PRODUCT IS PERFORMING AS EXPECTED. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS CONFIRMED BASED ON PHOTO RECEIVED OF A BOTTLE WITHOUT THE SENSOR. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO, PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. THE SPECIMEN MUST BE COLLECTED USING STERILE TECHNIQUES TO REDUCE THE CHANCE OF CONTAMINATION. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS-FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K970512. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BACTEC¿ MYCO/F LYTIC CULTURE VIALS THERE WAS ONE FALSE NEGATIVE RESULT. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BACTEC¿ MYCO/F LYTIC CULTURE VIALS THERE WAS ONE FALSE NEGATIVE RESULT. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1834885 BD BACTEC¿ MYCO/F LYTIC CULTURE VIALS SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 5057481 00382904422882

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown