PROTAPER SHAP. 19MM/SX*NOT CE*
Report
- Report Number
- 8031010-2025-00193
- Event Type
- Malfunction
- Date Received
- November 4, 2025
- Date of Event
- October 12, 2025
- Report Date
- December 22, 2025
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- EKS
- UDI-DI
- D716PTUHSX191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION RECEIVED THE BROKEN PART HAS BEEN RETRIEVED. THIS IS A FOLLOW UP REPORT FOR THIS ADDITIONAL INFORMATION. INVESTIGATION: THE INVESTIGATION RESULTS FOR THIS COMPLAINT ARE INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1891099). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE, OVERUSE (NUMBER OF USES NOT COMMUNICATED), EXCESSIVE WEAR, PATIENT CONDITION AND BEHAVIOR DURING TREATMENT OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE). IN ADDITION, THERE ARE FURTHER CIRCUMSTANCES ON THE CONDITIONS IN DENTISTRY (USED DISINFECTANTS, TRAINING/KNOWLEDGE STATUS), OR ANY OTHER ENVIRONMENTAL CONDITIONS, WHICH ARE UNKNOWN TO US AND MAY ALSO HAVE AN IMPACT ON THE REPORTED FAILURE MODE. THIS IS TO CORRECT AND REMOVE THE CODES THAT WERE INITIALLY REPORTED - REMOVING CODES FOR: HEALTH EFFECT - CLINICAL CODE; 3165. HEALTH EFFECT - IMPACT CODE: 4648. THE CORRECT CODES FOR THIS COMPLAINT ARE: HEALTH EFFECT - CLINICAL CODE: 4582. HEALTH EFFECT - IMPACT CODE: 2199. THIS IS A FOLLOW UP REPORT TO CORRECT THIS CODE.
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE LOT NUMBER WAS PROVIDED AND RETAINED-PRODUCT TESTING AND/OR DHR REVIEW ARE PLANNED. THE RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT IT IS REPORTED THAT A PROTAPER SHAP. 19MM/SX NOT CE FILE BROKE DURING USE. THE BROKEN PART REMAINS IN CANAL. THE OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 215779 | PROTAPER SHAP. 19MM/SX*NOT CE* | FILE, PULP CANAL, ENDODONTIC | EKS | MAILLEFER INSTRUMENTS HOLDING SARL | 1891099 | D716PTUHSX191 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |